Família ABC Arquivos Brasileiros de Cardiologia International Journal of Cardiovascular Sciences ABC Heart Failure & Cardiomyopathy ABC Imagem Cardiovascular
Português English
Família ABC
Português English
  • Sobre
    • Sobre a Família ABC
    • Regimento Interno
    • Conselho Editorial
    • Sociedade Brasileira de Cardiologia (SBC)
  • Edições
  • Diretrizes da SBC
  • Conteúdos
    • Vídeos
    • Podcasts
    • Notícias
  • Sobre
    • Sobre a Família ABC
    • Regimento Interno
    • Conselho Editorial
    • Sociedade Brasileira de Cardiologia (SBC)
  • Edições
  • Diretrizes da SBC
  • Conteúdos
    • Vídeos
    • Podcasts
    • Notícias

Família ABC

  • Família ABC
  • Arquivos Brasileiros de Cardiologia
  • International Journal of Cardiovascular Sciences
  • ABC Heart Failure & Cardiomyopathy
  • ABC Imagem Cardiovascular
Buscar em

Nenhum conteúdo encontrado.

Revistas
Ordenação
Tipos de artigo
Outros

  1. Início
  2. Palavra-chave: Sphygmomanometers
IJCS
Original Article

Evaluation of the Profile and Quality of Sphygmomanometers Available at a Health Education Institution in Belo Horizonte

Matheus Evangelista da Costa et al. Sarah Mattos Moraes Kleisson Antonio Pontes Maia

Abstract

Background

Arterial hypertension affects around 30% of the Brazilian adult population, showing a direct and progressive relationship with an increased risk of cardiovascular diseases (CVDs). In this context, adjustments in the calibration and physical constitution of sphygmomanometers are essential conditions for obtaining correct blood pressure (BP) measurements.

Objectives

Analyze the profile and quality of sphygmomanometers used in various sectors of a health education institution in Belo Horizonte, Minas Gerais.

Methods

The present study conducted a cross-sectional, observational, and non-interventional study to assess adherence to various quality parameters of the sphygmomanometers available in the outpatient clinic, emergency department, and ward of an educational institution in the municipality of Belo Horizonte.

Results

We analyzed 78 devices, in which high rates of adherence were identified for velcro/pins, 93.5% (73/78); bulbs/rubbers, 92.1% (70/76); valves, 93.4% (71/76); the seal of the National Institute of Metrology, Quality, and Technology (INMETRO), 97.4% (76/78); and the clock, 92.1% (70/76). However, these parameters showed no statistical significance. Institutions (hospital/outpatient clinic) recorded higher calibration rates, 75% (39/52, p<0.001), in accordance with the deadlines stipulated by INMETRO, and the cuff/equipment compatibility showed its highest value (52%) among students (32/61, p = 0.004).

Conclusion

Our study showed that 38.4% (30/78) of the devices did not presented some type of inadequacy, with outpatient-owned devices had the highest compliance rate (p = 0.015). These findings are worrisome, as they can lead to inaccurate BP measurements.

  • cardiovascular diseases
  • Equipment Failure
  • Hypertension
  • Sphygmomanometers
Ler artigo completo
Creative Commons Attribution (CC BY) Creative Commons license
11/03/2024 DOI: 10.36660/ijcs.20230175 Visualizações 2
IJCS
Original Article

Improvement in Semiconductivity On The Measurement Of Blood Pressure After An Educational Intervention In Health Professionals

Cynthia Kallás Bachur et al. Sarah Silva Candido Gerlia Bernardes Silveira Samantha Gurgel Oliveira Sousa … Joao Hercos Neto Eugenia Velludo Veiga

Abstract

Background

Measuring blood pressure is a simple method, but it is subject to errors.

Objective

to evaluate the theoretical and practical knowledge of the steps of blood pressure measurement in health professionals, before and after the educational intervention.

Methods

The theoretical knowledge questionnaire on indirect blood pressure measurement was used to assess theoretical knowledge; to assess practical knowledge, the simulation strategy was applied in a standardized clinical setting and environment. The assessments were reapplied after one month. For data analysis, descriptive statistics were used.

Results

30 health professionals from different categories; 19 of whom were males aged 41 ± 9.4 years and 11 were females aged 35 ± 9.5 years. Improvement was observed in most stages of theoretical and practical knowledge when compared to pre-and post-intervention, with an emphasis on the theoretical stages: “Position of the lower limbs” 2 (6.6%) x 16 (53.3%) and “Forceps with adequate position” 1 (3.3%) x 6 (20%). In the assessment of practical knowledge, it should be highlighted: “Do not speak during the measurement” 6 (20%) x 28 (93.3%) and in the “ideal size clamp” stage 0 (0%) x 5 (16.6 %).

Conclusion

The theoretical and practical knowledge on the stages of BP measurement by health professionals in this sample was insufficient. However, after the educational intervention, there was an improvement in the technique in most stages. (International Journal of Cardiovascular Sciences. 2020; [online].ahead print, PP.0-0)

  • Blood pressure
  • Health Personal
  • Hypertension
  • Inservice Training
  • Knowledge
  • Measument Equipment
  • questionnaires
  • Sphygmomanometers
  • Stethoscppes
Ler artigo completo
Creative Commons Attribution (CC BY) Creative Commons license
15/07/2020 DOI: 10.36660/ijcs.20190129 Visualizações 3
ABC
Original Article

The results of the Campaign for evaluating sphygmomanometers accuracy and their physical conditions

Décio Mion Jr et al. Angela M. G. Pierin Débora Cristina Alavarce José Henrique da Cunha Vasconcellos

OBJECTIVE: To evaluate the sphygmomanometers calibration accuracy and the physical conditions of the cuff-bladder, bulb, pump, and valve. METHODS: Sixty hundred and forty five aneroid sphygmomanometers were evaluated, 521 used in private practice and 124 used in hospitals. Aneroid manometers were tested against a properly calibrated mercury manometer and were considered calibrated when the error was <=3mm Hg. The physical conditions of the cuffs-bladder, bulb, pump, and valve were also evaluated. RESULTS: Of the aneroid sphygmomanometers tested, 51% of those used in private practice and 56% of those used in hospitals were found to be not accurately calibrated. Of these, the magnitude of inaccuracy ranged from 4 to 8mm Hg in 70% and 51% of the devices, respectively. The problems found in the cuffs – bladders, bulbs, pumps, and valves of the private practice and hospital devices were bladder damage (34% vs. 21%, respectively), holes/leaks in the bulbs (22% vs. 4%, respectively), and rubber aging (15% vs. 12%, respectively). Of the devices tested, 72% revealed at least one problem interfering with blood pressure measurement accuracy. CONCLUSION: Most of the manometers evaluated, whether used in private practice or in hospitals, were found to be inaccurate and unreliable, and their use may jeopardize the diagnosis and treatment of arterial hypertension.

  • blood pressure measurement
  • calibration accuracy
  • Sphygmomanometers
Ler artigo completo
Creative Commons Attribution (CC BY) Creative Commons license
05/09/2000 DOI: 10.1590/S0066-782X2000000100004 Visualizações 1

Como citar este artigo

EndNote RIS

Formato de Documento Eletrônico (Vancouver)

Costa ME, Moraes SM, Maia KAP. Evaluation of the Profile and Quality of Sphygmomanometers Available at a Health Education Institution in Belo Horizonte. International Journal of Cardiovascular Sciences 2024;37:e20230175.

Formato de Documento Eletrônico (ABNT)

Costa, Matheus Evangelista da; Moraes, Sarah Mattos; Maia, Kleisson Antonio Pontes. Evaluation of the Profile and Quality of Sphygmomanometers Available at a Health Education Institution in Belo Horizonte. International Journal of Cardiovascular Sciences, v. 37, e20230175, mar. 2024.

Formato de Documento Eletrônico (APA)

Costa, M. E. , Moraes, S. M., & Maia, K. A. P. (2024). Evaluation of the Profile and Quality of Sphygmomanometers Available at a Health Education Institution in Belo Horizonte. International Journal of Cardiovascular Sciences, 37, e20230175.

Formato de Documento Eletrônico (ISO)

Costa, Matheus Evangelista da and Moraes, Sarah Mattos and Maia, Kleisson Antonio Pontes. Evaluation of the Profile and Quality of Sphygmomanometers Available at a Health Education Institution in Belo Horizonte. International Journal of Cardiovascular Sciences [online]. 2024, vol. 37, [citado 2026-10-14], e20230175. Disponível em: <https://journals.cardiol.br/article/evaluation-of-the-profile-and-quality-of-sphygmomanometers-available-at-a-health-education-institution-in-belo-horizonte-2/>. ISSN 2359-4802

EndNote RIS

Formato de Documento Eletrônico (Vancouver)

Bachur CK, Candido SS, Silveira GB, Sousa SGO, Hercos Neto J, Veiga EV. Improvement in Semiconductivity On The Measurement Of Blood Pressure After An Educational Intervention In Health Professionals. International Journal of Cardiovascular Sciences 2020;34(1):1-7.

Formato de Documento Eletrônico (ABNT)

Bachur, Cynthia Kallás; Candido, Sarah Silva; Silveira, Gerlia Bernardes; Sousa, Samantha Gurgel Oliveira; Hercos Neto, Joao; Veiga, Eugenia Velludo. Improvement in Semiconductivity On The Measurement Of Blood Pressure After An Educational Intervention In Health Professionals. International Journal of Cardiovascular Sciences, v. 34, n. 1, p. 1-7, jul. 2020.

Formato de Documento Eletrônico (APA)

Bachur, C. K., Candido, S. S., Silveira, G. B., Sousa, S. G. O., Hercos Neto, J., & Veiga, E. V. (2020). Improvement in Semiconductivity On The Measurement Of Blood Pressure After An Educational Intervention In Health Professionals. International Journal of Cardiovascular Sciences, 34(1), 1-7.

Formato de Documento Eletrônico (ISO)

Bachur, Cynthia Kallás and Candido, Sarah Silva and Silveira, Gerlia Bernardes and Sousa, Samantha Gurgel Oliveira and Hercos Neto, Joao and Veiga, Eugenia Velludo. Improvement in Semiconductivity On The Measurement Of Blood Pressure After An Educational Intervention In Health Professionals. International Journal of Cardiovascular Sciences [online]. 2020, vol. 34, n. 1, [citado 2026-10-14], pp. 1-7. Disponível em: <https://journals.cardiol.br/article/improvement-in-semiconductivity-on-the-measurement-of-blood-pressure-after-an-educational-intervention-in-health-professionals-2/>. ISSN 2359-4802

EndNote RIS

Formato de Documento Eletrônico (Vancouver)

Mion Jr D, Pierin AMG, Alavarce DC, Vasconcellos JHC. The results of the Campaign for evaluating sphygmomanometers accuracy and their physical conditions. Arquivos Brasileiros de Cardiologia 2000;74(1):35-8.

Formato de Documento Eletrônico (ABNT)

Mion Jr, Décio; Pierin, Angela M. G.; Alavarce, Débora Cristina; Vasconcellos, José Henrique da Cunha. The results of the Campaign for evaluating sphygmomanometers accuracy and their physical conditions. Arquivos Brasileiros de Cardiologia, v. 74, n. 1, p. 35-38, set. 2000.

Formato de Documento Eletrônico (APA)

Mion Jr, D., Pierin, A. M. G., Alavarce, D. C., & Vasconcellos, J. H. C. (2000). The results of the Campaign for evaluating sphygmomanometers accuracy and their physical conditions. Arquivos Brasileiros de Cardiologia, 74(1), 35-38.

Formato de Documento Eletrônico (ISO)

Mion Jr, Décio and Pierin, Angela M. G. and Alavarce, Débora Cristina and Vasconcellos, José Henrique da Cunha. The results of the Campaign for evaluating sphygmomanometers accuracy and their physical conditions. Arquivos Brasileiros de Cardiologia [online]. 2000, vol. 74, n. 1, [citado 2026-10-14], pp. 35-38. Disponível em: <https://journals.cardiol.br/article/the-results-of-the-campaign-for-evaluating-sphygmomanometers-accuracy-and-their-physical-conditions-2/>. ISSN 0066-782X