<!DOCTYPE article PUBLIC "-//NLM//DTD JATS (Z39.96) Journal Publishing DTD v1.1 20151215//EN" "https://jats.nlm.nih.gov/publishing/1.1/JATS-journalpublishing1.dtd">
<article xmlns:xlink="http://www.w3.org/1999/xlink" specific-use="sps-1.9" dtd-version="1.1" xml:lang="pt" article-type="research-article">
  <front>
    <journal-meta>
      <journal-id journal-id-type="publisher-id">abc</journal-id>
      <journal-title-group>
        <journal-title>Arquivos Brasileiros de Cardiologia</journal-title>
        <abbrev-journal-title abbrev-type="publisher">Arq. Bras. Cardiol.</abbrev-journal-title>
      </journal-title-group>
      <issn pub-type="ppub">0066-782X</issn>
      <issn pub-type="epub">1678-4170</issn>
      <publisher>
        <publisher-name>Sociedade Brasileira de Cardiologia - SBC</publisher-name>
        <publisher-loc>S&#227;o Paulo, SP, Brazil</publisher-loc>
      </publisher>
    </journal-meta>
    <article-meta>
      <article-id pub-id-type="publisher-id" specific-use="scielo-v3">HgB3LzqkrDjXhrySz4pp5ky</article-id>
      <article-id pub-id-type="other">00003</article-id>
      <article-id pub-id-type="publisher-id" specific-use="scielo-v2">S0066-782X2009000100003</article-id>
      <article-id pub-id-type="publisher-id" specific-use="scielo-v1">S0066-782X(09)09200103</article-id>
      <article-id pub-id-type="doi">10.1590/S0066-782X2009000100003</article-id>
      <article-categories>
        <subj-group subj-group-type="heading">
          <subject>Artigos Originais</subject>
        </subj-group>
      </article-categories>
      <title-group>
        <article-title>Valida&#231;&#227;o do monitor de medida de press&#227;o arterial Omron HEM 742 em adolescentes</article-title>
      </title-group>
      <contrib-group>
        <contrib contrib-type="author">
          <name>
            <surname>Christofaro</surname>
            <given-names>Diego Giulliano Destro</given-names>
          </name>
          <xref ref-type="aff" rid="aff01"/>
        </contrib>
        <contrib contrib-type="author">
          <name>
            <surname>Fernandes</surname>
            <given-names>R&#244;mulo Ara&#250;jo</given-names>
          </name>
          <xref ref-type="aff" rid="aff02"/>
        </contrib>
        <contrib contrib-type="author">
          <name>
            <surname>Gerage</surname>
            <given-names>Aline Mendes</given-names>
          </name>
          <xref ref-type="aff" rid="aff01"/>
        </contrib>
        <contrib contrib-type="author">
          <name>
            <surname>Alves</surname>
            <given-names>Marcelo Jos&#233;</given-names>
          </name>
          <xref ref-type="aff" rid="aff02"/>
        </contrib>
        <contrib contrib-type="author">
          <name>
            <surname>Polito</surname>
            <given-names>Marcos Doederlein</given-names>
          </name>
          <xref ref-type="aff" rid="aff01"/>
        </contrib>
        <contrib contrib-type="author">
          <name>
            <surname>Oliveira</surname>
            <given-names>Arli Ramos de</given-names>
          </name>
          <xref ref-type="aff" rid="aff01"/>
        </contrib>
      </contrib-group>
      <aff id="aff01">
        <addr-line>
          <named-content content-type="city">Londrina</named-content>
          <named-content content-type="state">Paran&#225;</named-content>
        </addr-line>
        <institution content-type="orgname">Universidade Estadual de Londrina</institution>
        <country country="BR">Brazil</country>
      </aff>
      <aff id="aff02">
        <addr-line>
          <named-content content-type="city">S&#227;o Paulo</named-content>
          <named-content content-type="state">SP</named-content>
        </addr-line>
        <institution content-type="orgname">Universidade Estadual Paulista</institution>
        <country country="BR">Brasil</country>
      </aff>
      <pub-date date-type="collection" publication-format="electronic">
        <month>01</month>
        <year>2009</year>
      </pub-date>
      <volume>92</volume>
      <issue>1</issue>
      <fpage>10</fpage>
      <lpage>15</lpage>
      <history>
        <date date-type="rev-recd">
          <day>05</day>
          <month>05</month>
          <year>2008</year>
        </date>
        <date date-type="received">
          <day>29</day>
          <month>02</month>
          <year>2008</year>
        </date>
        <date date-type="accepted">
          <day>13</day>
          <month>05</month>
          <year>2008</year>
        </date>
      </history>
      <permissions>
        <license license-type="open-access" xlink:href="http://creativecommons.org/licenses/by-nc/4.0/" xml:lang="en">
          <license-p>This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.</license-p>
        </license>
      </permissions>
      <abstract>
        <p>FUNDAMENTO: A medida precisa da press&#227;o arterial &#233; fundamental para a investiga&#231;&#227;o cient&#237;fica ou decis&#227;o cl&#237;nica. Nesse sentido, &#233; importante verificar valores fornecidos por equipamentos eletr&#244;nicos. OBJETIVO: Validar o monitor Omron HEM 742 de medida de press&#227;o arterial em adolescentes, segundo os crit&#233;rios sugeridos pela British Hypertension Society. M&#201;TODOS: Participaram do estudo 150 adolescentes com idades entre 10 e 16 anos. O monitor autom&#225;tico Omron HEM 742 foi conectado em Y com equipamento auscultat&#243;rio de coluna de merc&#250;rio, e realizaram-se tr&#234;s avalia&#231;&#245;es simult&#226;neas, calculando-se as diferen&#231;as entre os dois equipamentos. Para verificar a rela&#231;&#227;o entre ambos, utilizaram-se o coeficiente de correla&#231;&#227;o intraclasse e a plotagem de Bland-Altman (concord&#226;ncia). A especificidade e a sensibilidade do aparelho foram determinadas pela curva ROC. RESULTADOS: A compara&#231;&#227;o entre as medidas acusou uma diferen&#231;a menor ou igual a 5 mmHg em 67,3% dos valores sist&#243;licos e 69,3% dos valores diast&#243;licos; uma diferen&#231;a menor ou igual a 10 mmHg ocorreu em 87,3% e 90,6% dos valores sist&#243;licos e diast&#243;licos, respectivamente; e uma diferen&#231;a menor ou igual a 15 mmHg em 96,6% dos valores sist&#243;licos e 97,3% dos diast&#243;licos. Esses resultados indicam grau A segundo o protocolo da British Hypertension Society. Observou-se ainda elevada concord&#226;ncia nos valores obtidos por meio do monitor autom&#225;tico, e verificou-se que esse equipamento &#233; capaz de identificar a presen&#231;a ou a aus&#234;ncia da press&#227;o arterial elevada. CONCLUS&#195;O: O monitor Omron HEM 742 mostrou-se v&#225;lido para medidas de press&#227;o arterial em adolescentes, conforme os crit&#233;rios sugeridos pela British Hypertension Society.</p>
      </abstract>
      <kwd-group xml:lang="pt" kwd-group-type="author-generated">
        <kwd>Press&#227;o arterial</kwd>
        <kwd>Monitores de press&#227;o arterial</kwd>
        <kwd>adolescente</kwd>
      </kwd-group>
      <pub-date publication-format="electronic" date-type="pub">
        <day>04</day>
        <month>02</month>
        <year>2009</year>
      </pub-date>
    </article-meta>
  </front>
  <body>
    <p>
      <bold>ARTIGO ORIGINAL</bold>
    </p>
    <p>
      <bold>Valida&#231;&#227;o do monitor de medida de press&#227;o arterial Omron HEM 742 em adolescentes </bold>
    </p>
    <p>
      <bold>Diego Giulliano Destro Christofaro<sup>I</sup>; R&#244;mulo Ara&#250;jo Fernandes<sup>II</sup>; Aline Mendes Gerage<sup>I</sup>; Marcelo Jos&#233; Alves<sup>II</sup>; Marcos Doederlein Polito<sup>I</sup>; Arli Ramos de Oliveira<sup>I</sup></bold>
    </p>
    <p><sup>I</sup>Universidade Estadual de Londrina, Londrina, PR - Brasil </p>
    <p><sup>II</sup>Universidade Estadual Paulista, S&#227;o Paulo, SP - Brasil</p>
    <p>
      <xref ref-type="corresp" rid="back">Correspond&#234;ncia</xref>
    </p>
    <p>
      <bold>RESUMO</bold>
    </p>
    <p><bold>FUNDAMENTO:</bold> A medida precisa da press&#227;o arterial &#233; fundamental para a investiga&#231;&#227;o cient&#237;fica ou decis&#227;o cl&#237;nica. Nesse sentido, &#233; importante verificar valores fornecidos por equipamentos eletr&#244;nicos. </p>
    <p><bold>OBJETIVO:</bold> Validar o monitor Omron HEM 742 de medida de press&#227;o arterial em adolescentes, segundo os crit&#233;rios sugeridos pela British Hypertension Society. </p>
    <p><bold>M&#201;TODOS:</bold> Participaram do estudo 150 adolescentes com idades entre 10 e 16 anos. O monitor autom&#225;tico Omron HEM 742 foi conectado em Y com equipamento auscultat&#243;rio de coluna de merc&#250;rio, e realizaram-se tr&#234;s avalia&#231;&#245;es simult&#226;neas, calculando-se as diferen&#231;as entre os dois equipamentos. Para verificar a rela&#231;&#227;o entre ambos, utilizaram-se o coeficiente de correla&#231;&#227;o intraclasse e a plotagem de Bland-Altman (concord&#226;ncia). A especificidade e a sensibilidade do aparelho foram determinadas pela curva ROC. </p>
    <p><bold>RESULTADOS:</bold> A compara&#231;&#227;o entre as medidas acusou uma diferen&#231;a menor ou igual a 5 mmHg em 67,3% dos valores sist&#243;licos e 69,3% dos valores diast&#243;licos; uma diferen&#231;a menor ou igual a 10 mmHg ocorreu em 87,3% e 90,6% dos valores sist&#243;licos e diast&#243;licos, respectivamente; e uma diferen&#231;a menor ou igual a 15 mmHg em 96,6% dos valores sist&#243;licos e 97,3% dos diast&#243;licos. Esses resultados indicam grau A segundo o protocolo da <italic>British Hypertension Society</italic>. Observou-se ainda elevada concord&#226;ncia nos valores obtidos por meio do monitor autom&#225;tico, e verificou-se que esse equipamento &#233; capaz de identificar a presen&#231;a ou a aus&#234;ncia da press&#227;o arterial elevada. </p>
    <p><bold>CONCLUS&#195;O:</bold> O monitor Omron HEM 742 mostrou-se v&#225;lido para medidas de press&#227;o arterial em adolescentes, conforme os crit&#233;rios sugeridos pela <italic>British Hypertension Society</italic></p>
    <p><bold>Palavras-chave:</bold> Press&#227;o arterial, Monitores de press&#227;o arterial/utiliza&#231;&#227;o, adolescente.</p>
    <p>
      <bold>Introdu&#231;&#227;o</bold>
    </p>
    <p>A preval&#234;ncia da press&#227;o arterial (PA) elevada tem crescido acentuadamente em homens e mulheres adultos de v&#225;rios pa&#237;ses<sup>1</sup>, sendo motivo de preocupa&#231;&#227;o por profissionais da sa&#250;de, uma vez que essa disfun&#231;&#227;o &#233; considerada um dos principais fatores de risco para o desenvolvimento de doen&#231;as cardiovasculares<sup>2,3</sup>. No entanto, dados nacionais mostram que crian&#231;as e adolescentes podem apresentar valores significativamente aumentados de PA<sup>4,5</sup>, o que tamb&#233;m &#233; preocupante levando-se em considera&#231;&#227;o que as condi&#231;&#245;es de risco durante a inf&#226;ncia e a adolesc&#234;ncia tendem a se expressar na vida adulta<sup>6</sup>.</p>
    <p>Medidas precisas e fidedignas de PA, seja na pr&#225;tica cl&#237;nica seja de forma experimental, tornam-se um procedimento imprescind&#237;vel para auxiliar no diagn&#243;stico ou na interpreta&#231;&#227;o de dados. Nesse sentido, a literatura aponta diversas formas de mensura&#231;&#227;o da PA<sup>7</sup>. Dentre esses m&#233;todos, o que tem sido mais comumente utilizado &#233; o m&#233;todo auscultat&#243;rio, seja por esfigmoman&#244;metros aner&#243;ides seja de coluna de merc&#250;rio. Por&#233;m, a acur&#225;cia desse m&#233;todo pode ser comprometida por fatores relativos ao observador.</p>
    <p>Assim, uma alternativa para a aferi&#231;&#227;o da PA &#233; a utiliza&#231;&#227;o de aparelhos eletr&#244;nicos. Atualmente, tem-se verificado um n&#250;mero crescente desses aparelhos no mercado, considerando que, al&#233;m de possu&#237;rem pre&#231;os acess&#237;veis, s&#227;o de f&#225;cil manipula&#231;&#227;o. Contudo, &#233; importante que esses equipamentos sejam avaliados de acordo com as normas de valida&#231;&#227;o exigidas por entidades internacionais, como a da <italic>British Hypertension Society</italic> (BHS)<sup>8</sup>. Os dados dispon&#237;veis na literatura quanto &#224; validade de aparelhos autom&#225;ticos utilizaram principalmente amostra de indiv&#237;duos adultos. Os experimentos utilizando crian&#231;as e adolescentes s&#227;o em menor n&#250;mero e, por isso, sugerem maiores investiga&#231;&#245;es.</p>
    <p>O presente estudo teve, portanto, como objetivo validar o aparelho oscilom&#233;trico autom&#225;tico Omron HEM 742 de medida de PA em crian&#231;as e adolescentes de acordo com o protocolo proposto pela BHS e avaliar a sua efici&#234;ncia na classifica&#231;&#227;o dos indiv&#237;duos dessa faixa et&#225;ria em rela&#231;&#227;o aos valores de PA de repouso.</p>
    <p>
      <bold>M&#233;todos</bold>
    </p>
    <p>
      <bold>Amostra</bold>
    </p>
    <p>As coletas foram efetuadas em uma institui&#231;&#227;o da rede estadual de ensino. A unidade escolar foi selecionada por conveni&#234;ncia por preencher todos os requisitos preestabelecidos pelos pesquisadores:</p>
    <disp-quote>
      <p>a) proximidade &#224; institui&#231;&#227;o de ensino superior;</p>
      <p>b) instala&#231;&#245;es adequadas para as aferi&#231;&#245;es;</p>
      <p>c) autoriza&#231;&#227;o do diretor respons&#225;vel pela unidade escolar;</p>
      <p>d) atender &#224; faixa et&#225;ria previamente definida pelos pesquisadores.</p>
    </disp-quote>
    <p>Da amostra analisada, foram exclu&#237;dos todos os sujeitos que, por declara&#231;&#227;o do respons&#225;vel, possu&#237;am arritmias ou fibrila&#231;&#227;o atrial, fizeram ingest&#227;o de cafe&#237;na 30 minutos antes da avalia&#231;&#227;o ou utilizavam f&#225;rmacos de a&#231;&#227;o crono ou inotr&#243;pica. Todos os sujeitos da amostra informaram n&#227;o utilizar tabaco e n&#227;o realizaram atividades f&#237;sicas no dia da aferi&#231;&#227;o da press&#227;o arterial. Assim, a amostra foi composta por 150 adolescentes (77 do sexo masculino e 73 do feminino).</p>
    <p>Todos os respons&#225;veis pelos sujeitos assinaram um termo de consentimento livre e esclarecido concordando em participar do estudo, de acordo com as instru&#231;&#245;es contidas na Resolu&#231;&#227;o n&#186; 196/96 do Conselho Nacional de Sa&#250;de para estudos com seres humanos. O estudo foi aprovado pelo Comit&#234; de &#201;tica da Institui&#231;&#227;o de Ensino Superior.</p>
    <p>
      <bold>Medidas antropom&#233;tricas</bold>
    </p>
    <p>A idade cronol&#243;gica dos adolescentes foi determinada em forma centesimal, utilizando a data de nascimento e o dia da avalia&#231;&#227;o. Foram realizadas medidas de massa corporal e de estatura para a caracteriza&#231;&#227;o da amostra. Para tanto, utilizou-se uma balan&#231;a port&#225;til digital (gradua&#231;&#227;o: 100 g e capacidade m&#225;xima: 150 kg) e um estadi&#244;metro de metal fixado &#224; parede (precis&#227;o: 0,1 cm e extens&#227;o m&#225;xima: dois metros), sendo todas as medidas efetuadas com os indiv&#237;duos descal&#231;os e vestindo roupas leves, de acordo com o protocolo sugerido por Gordon e cols.<sup>9</sup>. A partir dessas informa&#231;&#245;es, o &#237;ndice de massa corporal (IMC) foi calculado, sendo os valores expressos em quilogramas por metro quadrado (kg/m<sup>2</sup>).</p>
    <p>
      <bold>Medidas de press&#227;o arterial</bold>
    </p>
    <p>Para a mensura&#231;&#227;o dos valores de PA entre os adolescentes, utilizaram-se dois tipos de manguitos de acordo com a circunfer&#234;ncia do bra&#231;o (6 mm x 12 mm (tamanho infantil) e 9 mm x 18 mm (para adolescentes de 14-16 anos e para aqueles com idade inferior mas que possu&#237;ssem uma estrutura f&#237;sica maior), seguindo as recomenda&#231;&#245;es da Associa&#231;&#227;o Americana do Cora&#231;&#227;o<sup>7</sup>. Todas as avalia&#231;&#245;es da medida da press&#227;o arterial foram realizadas no bra&#231;o direito de cada indiv&#237;duo. O equipamento a ser testado foi o Omron HEM 742, que consiste em um aparelho eletr&#244;nico e digital de medida de PA de bra&#231;o, com infla&#231;&#227;o e defla&#231;&#227;o autom&#225;tica do ar. O m&#233;todo de medida desse aparelho &#233; o oscilom&#233;trico, com varia&#231;&#227;o da press&#227;o de 0-280 mmHg.</p>
    <p>O aparelho oscilom&#233;trico foi conectado em Y com o esfigmoman&#244;metro de coluna de merc&#250;rio. O mecanismo de infla&#231;&#227;o do aparelho foi ativado para realiza&#231;&#227;o de tr&#234;s medidas consecutivas com intervalo de 2 minutos entre elas. No final de cada medida, um observador registrava os valores da press&#227;o arterial sist&#243;lica (PAS) e diast&#243;lica (PAD), identificados na coluna de merc&#250;rio, sem ter conhecimento dos valores registrados no aparelho autom&#225;tico reportados por outro observador.</p>
    <p>Antes de serem avaliados, os indiv&#237;duos permaneceram 5 minutos sentados em repouso com o tronco apoiado em uma cadeira e com o bra&#231;o relaxado, e a coluna de merc&#250;rio e o aparelho autom&#225;tico foram devidamente calibrados antes de serem executadas as aferi&#231;&#245;es. O tempo demandado para a avalia&#231;&#227;o total em cada sujeito da amostra (avalia&#231;&#227;o antropom&#233;trica e da press&#227;o arterial) foi aproximadamente de 10 minutos.</p>
    <p>Ao todo foram realizadas simultaneamente 450 avalia&#231;&#245;es com ambos os instrumentos de medidas da PA. A classifica&#231;&#227;o da varia&#231;&#227;o dos valores de PA fornecidos pelo aparelho foi feita considerando as diferen&#231;as dos registros das leituras entre o man&#244;metro da coluna de merc&#250;rio e o aparelho autom&#225;tico, conforme os procedimentos descritos pela BHS. O protocolo da BHS indica que os aparelhos devem alcan&#231;ar no m&#237;nimo o grau B para PAS e PAD para serem instrumentos v&#225;lidos. Os respectivos graus e seus valores est&#227;o descritos no <xref ref-type="fig" rid="q1">quadro 1</xref>.</p>
    <fig id="q1">
      <graphic xlink:href="1678-4170-abc-92-01-0010-00003-gf01.jpg"/></fig>
    <p>
      <bold>An&#225;lise estat&#237;stica</bold>
    </p>
    <p>Inicialmente, foi aplicado o teste de Kolmogorov-Smirnov para verificar a distribui&#231;&#227;o dos dados. A partir da&#237;, realizou-se a estat&#237;stica descritiva utilizando valores de m&#233;dia e desvio padr&#227;o (DP), uma vez que todas as vari&#225;veis analisadas se enquadraram no modelo Gaussiano de distribui&#231;&#227;o. A an&#225;lise de vari&#226;ncia para medidas repetidas (com corre&#231;&#227;o de Greenhouse-Geisser quando violado o pressuposto da esfericidade) e o teste t de Student para amostras pareadas estabeleceram compara&#231;&#245;es entre os valores m&#233;dios de PA encontrados entre os dois aparelhos. Para verificar a rela&#231;&#227;o entre os valores de PA fornecidos pelo aparelho autom&#225;tico e a coluna de merc&#250;rio, foi realizado o coeficiente de correla&#231;&#227;o intraclasse (CCI). As medidas de concord&#226;ncia foram testadas por meio da plotagem de Bland-Altman (valores num&#233;ricos), e os par&#226;metros fornecidos pela curva ROC (sensibilidade, especificidade e &#225;rea sob a curva (AUC)) indicaram o poder de diagn&#243;stico de valores elevados de press&#227;o arterial do aparelho oscilom&#233;trico. Valores de signific&#226;ncia inferiores a 5% foram considerados relevantes do ponto de vista estat&#237;stico, e os programas utilizados foram o MedCalc e SPSS vers&#227;o 13.0.</p>
    <p>
      <bold>Resultados</bold>
    </p>
    <p>Na <xref ref-type="table" rid="tab1">tabela 1</xref>, s&#227;o apresentadas as caracter&#237;sticas gerais da amostra agrupadas de acordo com o g&#234;nero. N&#227;o foram observadas diferen&#231;as estat&#237;sticas entre as vari&#225;veis analisadas.</p>
    <table-wrap id="tab1">
      <graphic xlink:href="1678-4170-abc-92-01-0010-00003-gt01.jpg"/></table-wrap>
    <p>Os resultados expostos na <xref ref-type="table" rid="tab2">tabela 2</xref> demonstram, em valores percentuais, as diferen&#231;as entre as medidas obtidas pelo aparelho autom&#225;tico e a coluna de merc&#250;rio. Conforme os resultados encontrados, o aparelho testado obteve grau A tanto para a PAS quanto para a PAD, segundo os crit&#233;rios estabelecidos pela BHS.</p>
    <table-wrap id="tab2">
      <graphic xlink:href="1678-4170-abc-92-01-0010-00003-gt02.jpg"/></table-wrap>
    <p>Na <xref ref-type="table" rid="t3">tabela, 3</xref> s&#227;o apresentadas as compara&#231;&#245;es e os CCI para os valores de PAS e PAD. Entre os aparelhos utilizados e em todas as tr&#234;s aferi&#231;&#245;es, houve diferen&#231;as significativas para os valores de PAS. Entretanto, para a PAD houve diferen&#231;a apenas na primeira aferi&#231;&#227;o. Em ambos os aparelhos, n&#227;o foram observadas diferen&#231;as estat&#237;sticas para os valores de PAS e PAD entre as tr&#234;s aferi&#231;&#245;es realizadas. Em rela&#231;&#227;o ao g&#234;nero, os valores de correla&#231;&#227;o para PAS observados entre os aparelhos oscilaram de moderado a alto (CCI = 0,74 e CCI = 0,88). J&#225; para a PAD, foram observados valores moderados (CCI = 0,50 e CCI = 0,79). Em ambas as circunst&#226;ncias, n&#227;o foram observadas diferen&#231;as entre os g&#234;neros que comprometessem a utiliza&#231;&#227;o do aparelho.</p>
    <p content-type="asset">
      <table-wrap id="t3">
        <graphic xlink:href="1678-4170-abc-92-01-0010-00003-gt03.jpg"/></table-wrap>
    </p>
    <p>A plotagem de Bland-Altman (<xref ref-type="fig" rid="f1">fig. 1</xref>) foi utilizada como indicador de concord&#226;ncia para os valores de PA reportados por ambos os aparelhos. Para os valores de PAS, nas 450 aferi&#231;&#245;es, verificou-se que apenas 2,2% das observa&#231;&#245;es situaram-se fora do intervalo de confian&#231;a de 95% estipulado para a m&#233;dia das diferen&#231;as observadas. J&#225; para a PAD, somente 2,4% das observa&#231;&#245;es situaram-se fora do mesmo intervalo de confian&#231;a.</p>
    <p content-type="asset">
      <fig id="f1">
        <graphic xlink:href="1678-4170-abc-92-01-0010-00003-gf02.jpg"/></fig>
    </p>
    <p>Nas 450 aferi&#231;&#245;es, ap&#243;s a classifica&#231;&#227;o dos indiv&#237;duos segundo seu comportamento hemodin&#226;mico em repouso (normotens&#227;o/PA elevada), foi testada a efici&#234;ncia do aparelho autom&#225;tico em diagnosticar a presen&#231;a da PA elevada (<xref ref-type="table" rid="tab4">tab. 4</xref>).</p>
    <table-wrap id="tab4">
      <graphic xlink:href="1678-4170-abc-92-01-0010-00003-gt04.jpg"/></table-wrap>
    <p>A <xref ref-type="table" rid="tab4">tabela 4</xref> indica que, nas tr&#234;s aferi&#231;&#245;es, foram observados elevados valores de sensibilidade (indica&#231;&#227;o correta da presen&#231;a da vari&#225;vel desfecho), de especificidade (indica&#231;&#227;o correta da aus&#234;ncia da vari&#225;vel desfecho) e, conseq&#252;entemente, de AUC (coeficiente geral do diagn&#243;stico, derivado da intera&#231;&#227;o entre sensibilidade e especificidade).</p>
    <p>
      <bold>Discuss&#227;o</bold>
    </p>
    <p>O prop&#243;sito deste estudo foi validar o aparelho oscilom&#233;trico autom&#225;tico Omron HEM 742 de medida de PA em crian&#231;as e adolescentes de acordo com o protocolo proposto pela BHS e avaliar a sua efici&#234;ncia na classifica&#231;&#227;o dos indiv&#237;duos dessa faixa et&#225;ria em rela&#231;&#227;o aos valores de PA de repouso, considerando que este aparelho foi validado apenas em indiv&#237;duos adultos<sup>10</sup> e que a maioria dos estudos envolvendo a valida&#231;&#227;o de outros aparelhos autom&#225;ticos para a aferi&#231;&#227;o da PA foi realizada com popula&#231;&#245;es adultas<sup>11-14</sup>. Al&#233;m disso, com aumento observado na ocorr&#234;ncia de valores elevados de PA de repouso entre crian&#231;as e adolescentes<sup>4,5</sup>, faz-se necess&#225;ria tamb&#233;m a realiza&#231;&#227;o desses estudos com popula&#231;&#245;es mais jovens, no sentido de serem criadas estrat&#233;gias que facilitem a medida de PA e sua interpreta&#231;&#227;o nessa popula&#231;&#227;o, tanto na pr&#225;tica cl&#237;nica quanto em pesquisas.</p>
    <p>Ressalta-se, dessa forma, que aparelhos de insufla&#231;&#227;o e defla&#231;&#227;o autom&#225;tica s&#227;o leves, port&#225;teis e de f&#225;cil aplica&#231;&#227;o e interpreta&#231;&#227;o, diminuindo a ocorr&#234;ncia de erros de leitura pelo avaliador, o que &#233; relativamente observado pelo m&#233;todo auscultat&#243;rio. Portanto, a utiliza&#231;&#227;o de aparelhos autom&#225;ticos parece ser uma estrat&#233;gia interessante para a mensura&#231;&#227;o da PA, principalmente quando se trata de um maior controle residencial dos n&#237;veis press&#243;ricos por parte de popula&#231;&#245;es hipertensas.</p>
    <p>Para que um aparelho seja validado para a utiliza&#231;&#227;o em popula&#231;&#245;es especiais, a BHS<sup>8</sup> sugere que ele j&#225; seja validado em adultos, como &#233; o caso do aparelho Omron HEM 742, validado por Coleman e cols.<sup>10</sup>, e que diferentes valores de PA sejam avaliados, variando de 100 mmHg a 240 mmHg para PAS e de 60 mmHg a 120 mmHg para PAD. Essa, todavia, foi uma limita&#231;&#227;o do estudo, uma vez que os valores de PAS oscilaram de 98 mmHg a 135 mmHg e de 53 mmHg a 86 mmHg para a PAD. Entretanto, essa varia&#231;&#227;o encontrada na presente investiga&#231;&#227;o &#233; semelhante &#224; observada em outros estudos de valida&#231;&#227;o de aparelhos para a mensura&#231;&#227;o da PA em popula&#231;&#245;es jovens<sup>15,16</sup>, nos quais valores superiores a 140 mmHg para PAS e 90 mmHg para PAD n&#227;o foram encontrados.</p>
    <p>Isso ocorre porque os &#237;ndices de preval&#234;ncia de hipertens&#227;o arterial nessa faixa et&#225;ria s&#227;o menores<sup>17,18</sup> quando comparados &#224; popula&#231;&#227;o adulta<sup>18-22</sup>. Al&#233;m do mais, ressalta-se que o diagn&#243;stico da hipertens&#227;o arterial na popula&#231;&#227;o jovem ocorre por meio de percentis, considerando o g&#234;nero, a idade e a estatura, diferentemente dos adultos que possuem valores de PAS e PAD preestabelecidos.</p>
    <p>No presente estudo, em todas as tr&#234;s aferi&#231;&#245;es realizadas na amostra, os valores de PAS fornecidos pelo aparelho testado superestimaram significativamente os valores obtidos pelo instrumento de coluna de merc&#250;rio. Esses maiores valores fornecidos pelo aparelho oscilom&#233;trico podem ser origin&#225;rios de medidas mais baixas produzidas pelo avaliador do m&#233;todo auscultat&#243;rio e/ou ser realmente valores mais altos produzidos pelo aparelho teste<sup>23</sup>. Por&#233;m, uma explica&#231;&#227;o definitiva para esse comportamento n&#227;o est&#225; totalmente definida na literatura. Em rela&#231;&#227;o &#224; PAD, somente foi encontrada diferen&#231;a significativa na primeira aferi&#231;&#227;o. Apesar de tais diferen&#231;as, os valores de PAS e PAD apresentados pelo aparelho Omron HEM 742 e pelo aparelho de coluna de merc&#250;rio n&#227;o oscilaram de maneira significativa entre as tr&#234;s aferi&#231;&#245;es quando comparados consigo mesmos, sugerindo boa reprodutibilidade dos dados em ambos os instrumentos.</p>
    <p>O CCI apresentou-se significativo e com valores moderados/elevados em todas as aferi&#231;&#245;es, tanto para PAS como para PAD, indicando boa rela&#231;&#227;o entre os valores observados para ambos os instrumentos. Ou seja, quanto maior for o valor reportado por um aparelho, maior ser&#225; o valor encontrado pelo outro aparelho para o mesmo indiv&#237;duo. Contudo, tal procedimento estat&#237;stico &#233; apenas uma medida de rela&#231;&#227;o, n&#227;o indicando a concord&#226;ncia entre esses equipamentos e, conseq&#252;entemente, n&#227;o detectando poss&#237;veis varia&#231;&#245;es extremas entre eles. Tal limita&#231;&#227;o &#233; suplantada pela utiliza&#231;&#227;o da plotagem de Bland-Altman que, por sua vez, indica a concord&#226;ncia entre dados num&#233;ricos. No presente estudo, foi observada uma baixa discord&#226;ncia entre os valores de PA encontrados por ambos os aparelhos. Esse fato confirmou-se quando visto que os crit&#233;rios propostos pela BSH<sup>8</sup> foram atendidos, atingindo-se, inclusive, a gradua&#231;&#227;o mais elevada.</p>
    <p>Tais indicativos positivos apresentados pelo CCI e pela plotagem de Bland-Altman s&#227;o informa&#231;&#245;es importantes sobre a confiabilidade da utiliza&#231;&#227;o do instrumento em quest&#227;o. Por&#233;m, trata-se de valores num&#233;ricos e, por isso, n&#227;o s&#227;o capazes de indicar a efici&#234;ncia do instrumento na pr&#225;tica cl&#237;nica, mais especificamente na detec&#231;&#227;o de valores elevados de PA, sendo necess&#225;rio empregar a curva ROC para esse fim. Assim, os par&#226;metros fornecidos pela curva ROC indicaram que o aparelho Omron HEM 742 apresentou elevada capacidade de detectar tanto a presen&#231;a (sensibilidade) quanto a aus&#234;ncia (especificidade) de valores elevados de PA na amostra investigada.</p>
    <p>
      <bold>Conclus&#227;o</bold>
    </p>
    <p>O aparelho Omron HEM 742 apresenta boa confiabilidade para uso na avalia&#231;&#227;o de adolescentes, uma vez que foi validado de acordo com as recomenda&#231;&#245;es internacionais da BHS<sup>8</sup>, atingindo grau A para as an&#225;lises da PAS e PAD, al&#233;m de ter apresentado boa sensibilidade e especificidade no diagn&#243;stico em rela&#231;&#227;o aos n&#237;veis press&#243;ricos de indiv&#237;duos dessa faixa et&#225;ria.</p>
    <p>
      <bold>Potencial Conflito de Interesses</bold>
    </p>
    <p>Declaro n&#227;o haver conflito de interesses pertinentes.</p>
    <p>
      <bold>Fontes de Financiamento</bold>
    </p>
    <p>O presente estudo n&#227;o teve fontes de financiamento externas.</p>
    <p>
      <bold>Vincula&#231;&#227;o Acad&#234;mica</bold>
    </p>
    <p>Este artigo &#233; parte de disserta&#231;&#227;o de Mestrado de Diego Giulliano Destro Christofaro pela Universidade Estadual de Londrina.</p>
    <p>
      <fn id="back" fn-type="corresp">
        <p>Correspond&#234;ncia:</p>
        <p> Diego Giulliano Destro Christofaro </p>
        <p> Rua Belo Horizonte, 99, ap. 704 - Centro </p>
        <p> 86001-970 - Londrina - PR </p>
        <p> E-mail: </p>
        <p>
          <email xlink:href="diegochristofaro@yahoo.com.br">diegochristofaro@yahoo.com.br</email>
        </p>
      </fn>
    </p>
    <p>Artigo recebido em 29/02/08; revisado recebido em 05/05/08; aceito em 13/05/08</p>
  </body>
  <back>
    <ref-list>
      <ref id="B1">
        <mixed-citation>1. Cooper RS, Wolf-Maier    K, Luke A, Adeyemo A, Banegas JR, Forrester T, et al. An international comparative    study of blood pressure in populations of European vs. African descent. BMC    Med.2005;3:1-8.</mixed-citation>
        <element-citation publication-type="journal">
          <article-title>An international comparative study of blood pressure in populations of European vs. African descent</article-title>
          <source>BMC Med</source>
          <date>
            <year>2005</year>
          </date>
          <fpage>1</fpage>
          <lpage>8</lpage>
          <volume>3</volume>
          <person-group person-group-type="author">
            <name>
              <surname>Cooper</surname>
              <given-names>RS</given-names>
            </name>
            <name>
              <surname>Wolf-Maier</surname>
              <given-names>K</given-names>
            </name>
            <name>
              <surname>Luke</surname>
              <given-names>A</given-names>
            </name>
            <name>
              <surname>Adeyemo</surname>
              <given-names>A</given-names>
            </name>
            <name>
              <surname>Banegas</surname>
              <given-names>JR</given-names>
            </name>
            <name>
              <surname>Forrester</surname>
              <given-names>T</given-names>
            </name>
          </person-group>
        </element-citation>
      </ref>
      <ref id="B2">
        <mixed-citation>2. Chobanian AV,    Bakris GL, Black HR, Cushman WC, Green LA, Izzo JL Jr, et al. VII JNC. Seventh    Report of the Joint National Committee on Prevention, Detection, Evaluation    and Treatment of High Blood Pressure, Hypertension. 2003; 42 (6): 1206-52.</mixed-citation>
        <element-citation publication-type="journal">
          <article-title>VII JNC</article-title>
          <source>Seventh Report of the Joint National Committee on Prevention, Detection, Evaluation and Treatment of High Blood Pressure, Hypertension</source>
          <date>
            <year>2003</year>
          </date>
          <fpage>1206</fpage>
          <lpage>52</lpage>
          <issue>6</issue>
          <volume>42</volume>
          <person-group person-group-type="author">
            <name>
              <surname>Chobanian</surname>
              <given-names>AV</given-names>
            </name>
            <name>
              <surname>Bakris</surname>
              <given-names>GL</given-names>
            </name>
            <name>
              <surname>Black</surname>
              <given-names>HR</given-names>
            </name>
            <name>
              <surname>Cushman</surname>
              <given-names>WC</given-names>
            </name>
            <name>
              <surname>Green</surname>
              <given-names>LA</given-names>
            </name>
            <name>
              <surname>Izzo Jr</surname>
              <given-names>JL</given-names>
            </name>
          </person-group>
        </element-citation>
      </ref>
      <ref id="B3">
        <mixed-citation>3. Whelton SP,    Chin A, Xin X, He J. Effect of aerobic exercise on blood pressure: a meta-analysis    of randomized, controlled trials. Ann Intern Med. 2002; 136 (7): 493-503.</mixed-citation>
        <element-citation publication-type="journal">
          <article-title>Effect of aerobic exercise on blood pressure: a meta-analysis of randomized, controlled trials</article-title>
          <source>Ann Intern Med</source>
          <date>
            <year>2002</year>
          </date>
          <fpage>493</fpage>
          <lpage>503</lpage>
          <issue>7</issue>
          <volume>136</volume>
          <person-group person-group-type="author">
            <name>
              <surname>Whelton</surname>
              <given-names>SP</given-names>
            </name>
            <name>
              <surname>Chin</surname>
              <given-names>A</given-names>
            </name>
            <name>
              <surname>Xin</surname>
              <given-names>X</given-names>
            </name>
            <name>
              <surname>He</surname>
              <given-names>J</given-names>
            </name>
          </person-group>
        </element-citation>
      </ref>
      <ref id="B4">
        <mixed-citation>4. Nogueira PC,    da Costa RF, Cunha JS, Silvestrini L, Fisberg M. High arterial pressure in school    children in Santos-relationship to obesity. Rev Ass Med Bras. 2007; 53 (5):    426-32.</mixed-citation>
        <element-citation publication-type="journal">
          <article-title>High arterial pressure in school children in Santos-relationship to obesity</article-title>
          <source>Rev Ass Med Bras</source>
          <date>
            <year>2007</year>
          </date>
          <fpage>426</fpage>
          <lpage>32</lpage>
          <issue>5</issue>
          <volume>53</volume>
          <person-group person-group-type="author">
            <name>
              <surname>Nogueira</surname>
              <given-names>PC</given-names>
            </name>
            <name>
              <surname>da Costa</surname>
              <given-names>RF</given-names>
            </name>
            <name>
              <surname>Cunha</surname>
              <given-names>JS</given-names>
            </name>
            <name>
              <surname>Silvestrini</surname>
              <given-names>L</given-names>
            </name>
            <name>
              <surname>Fisberg</surname>
              <given-names>M</given-names>
            </name>
          </person-group>
        </element-citation>
      </ref>
      <ref id="B5">
        <mixed-citation>5. Moura AA, Silva    MAM, Ferraz MRMT, Rivera IR. Preval&#234;ncia de press&#227;o arterial elevada    em escolares e adolescentes de Macei&#243;. J Pediatr. 2004; 80 (1): 35-40.</mixed-citation>
        <element-citation publication-type="journal">
          <article-title>Preval&#234;ncia de press&#227;o arterial elevada em escolares e adolescentes de Macei&#243;</article-title>
          <source>J Pediatr</source>
          <date>
            <year>2004</year>
          </date>
          <fpage>35</fpage>
          <lpage>40</lpage>
          <issue>1</issue>
          <volume>80</volume>
          <person-group person-group-type="author">
            <name>
              <surname>Moura</surname>
              <given-names>AA</given-names>
            </name>
            <name>
              <surname>Silva</surname>
              <given-names>MAM</given-names>
            </name>
            <name>
              <surname>Ferraz</surname>
              <given-names>MRMT</given-names>
            </name>
            <name>
              <surname>Rivera</surname>
              <given-names>IR</given-names>
            </name>
          </person-group>
        </element-citation>
      </ref>
      <ref id="B6">
        <mixed-citation>6. Lauer RM, Clarke    WR. Childhood risk factors for high adult blood pressure: the Muscatine Study.    Pediatrics. 1989; 84 (4): 633-41.</mixed-citation>
        <element-citation publication-type="journal">
          <article-title>Childhood risk factors for high adult blood pressure: the Muscatine Study</article-title>
          <source>Pediatrics</source>
          <date>
            <year>1989</year>
          </date>
          <fpage>633</fpage>
          <lpage>41</lpage>
          <issue>4</issue>
          <volume>84</volume>
          <person-group person-group-type="author">
            <name>
              <surname>Lauer</surname>
              <given-names>RM</given-names>
            </name>
            <name>
              <surname>Clarke</surname>
              <given-names>WR</given-names>
            </name>
          </person-group>
        </element-citation>
      </ref>
      <ref id="B7">
        <mixed-citation>7. Pickering TG,    Hall JE, Appel LJ, Falkner BE, Graves J, Hill MN, et al. Subcommittee of Professional    and Public Education of the American Heart Association Council on High Blood    Pressure Research. Recommendations for blood pressure measurement in humans    and experimental animals: Part 1: blood pressure measurement in humans: a statement    for professionals from the Subcommittee of Professional and Public Education    of the American Heart Association Council on High Blood Pressure Research. Hypertension.    2005; 45 (5): 142-61.</mixed-citation>
        <element-citation publication-type="journal">
          <article-title>Subcommittee of Professional and Public Education of the American Heart Association Council on High Blood Pressure Research: Recommendations for blood pressure measurement in humans and experimental animals: Part 1: blood pressure measurement in humans: a statement for professionals from the Subcommittee of Professional and Public Education of the American Heart Association Council on High Blood Pressure Research</article-title>
          <source>Hypertension</source>
          <date>
            <year>2005</year>
          </date>
          <fpage>142</fpage>
          <lpage>61</lpage>
          <issue>5</issue>
          <volume>45</volume>
          <person-group person-group-type="author">
            <name>
              <surname>Pickering</surname>
              <given-names>TG</given-names>
            </name>
            <name>
              <surname>Hall</surname>
              <given-names>JE</given-names>
            </name>
            <name>
              <surname>Appel</surname>
              <given-names>LJ</given-names>
            </name>
            <name>
              <surname>Falkner</surname>
              <given-names>BE</given-names>
            </name>
            <name>
              <surname>Graves</surname>
              <given-names>J</given-names>
            </name>
            <name>
              <surname>Hill</surname>
              <given-names>MN</given-names>
            </name>
          </person-group>
        </element-citation>
      </ref>
      <ref id="B8">
        <mixed-citation>8. O&#180;Brien    E, Petrie J, Littler W, de Swiet M, Padfield Pl, O'Malley K, et al. The British    Hypertension society protocol for the evaluation of automated and semi-automated    blood pressure measuring devices with special reference to ambulatory systems.    J Hypertens. 1990; 8 (7): 607-19.</mixed-citation>
        <element-citation publication-type="journal">
          <article-title>The British Hypertension society protocol for the evaluation of automated and semi-automated blood pressure measuring devices with special reference to ambulatory systems</article-title>
          <source>J Hypertens</source>
          <date>
            <year>1990</year>
          </date>
          <fpage>607</fpage>
          <lpage>19</lpage>
          <issue>7</issue>
          <volume>8</volume>
          <person-group person-group-type="author">
            <name>
              <surname>O&#180;Brien</surname>
              <given-names>E</given-names>
            </name>
            <name>
              <surname>Petrie</surname>
              <given-names>J</given-names>
            </name>
            <name>
              <surname>Littler</surname>
              <given-names>W</given-names>
            </name>
            <name>
              <surname>de Swiet</surname>
              <given-names>M</given-names>
            </name>
            <name>
              <surname>Padfield</surname>
              <given-names>Pl</given-names>
            </name>
            <name>
              <surname>O'Malley</surname>
              <given-names>K</given-names>
            </name>
          </person-group>
        </element-citation>
      </ref>
      <ref id="B9">
        <mixed-citation>9. Gordon CC, Chumlea    WC, Roche AF. Stature, recumbent length and weight. In: Lohman TG, Roche AF,    Martorel R, editors. Anthropometric standardization reference manual. Champaign:    Human Kinetics Books; 1988. p. 3-8.</mixed-citation>
        <element-citation publication-type="book">
          <source>Anthropometric standardization reference manual</source>
          <date>
            <year>1988</year>
          </date>
          <fpage>3</fpage>
          <lpage>8</lpage>
          <person-group person-group-type="author">
            <name>
              <surname>Gordon</surname>
              <given-names>CC</given-names>
            </name>
            <name>
              <surname>Chumlea</surname>
              <given-names>WC</given-names>
            </name>
            <name>
              <surname>Roche</surname>
              <given-names>AF</given-names>
            </name>
            <name>
              <surname>Lohman</surname>
              <given-names>TG</given-names>
            </name>
            <name>
              <surname>Roche</surname>
              <given-names>AF</given-names>
            </name>
            <name>
              <surname>Martorel</surname>
              <given-names>R</given-names>
            </name>
          </person-group>
        </element-citation>
      </ref>
      <ref id="B10">
        <mixed-citation>10. Coleman A,    Freeman P, Steel S, Shennan A. Validation of the Omron MX3 Plus oscilometric    blood pressure monitoring device according to the European Society of Hypertension    International protocol. Blood Press Monit.2005;10(3):165-8.</mixed-citation>
        <element-citation publication-type="journal">
          <article-title>Validation of the Omron MX3 Plus oscilometric blood pressure monitoring device according to the European Society of Hypertension International protocol</article-title>
          <source>Blood Press Monit</source>
          <date>
            <year>2005</year>
          </date>
          <fpage>165</fpage>
          <lpage>8</lpage>
          <issue>3</issue>
          <volume>10</volume>
          <person-group person-group-type="author">
            <name>
              <surname>Coleman</surname>
              <given-names>A</given-names>
            </name>
            <name>
              <surname>Freeman</surname>
              <given-names>P</given-names>
            </name>
            <name>
              <surname>Steel</surname>
              <given-names>S</given-names>
            </name>
            <name>
              <surname>Shennan</surname>
              <given-names>A</given-names>
            </name>
          </person-group>
        </element-citation>
      </ref>
      <ref id="B11">
        <mixed-citation>11. Altunkan S,    Ilman N, Kayat&#252;rk N, Altunkan E. Validation of the Omron M6 (HEM-7001-E)    upper-arm blood pressure measuring device according to the International Protocol    in adults and obese adults. Blood Press Monit. 2007; 12 (4): 219-25.</mixed-citation>
        <element-citation publication-type="journal">
          <article-title>Validation of the Omron M6 (HEM-7001-E) upper-arm blood pressure measuring device according to the International Protocol in adults and obese adults</article-title>
          <source>Blood Press Monit</source>
          <date>
            <year>2007</year>
          </date>
          <fpage>219</fpage>
          <lpage>25</lpage>
          <issue>4</issue>
          <volume>12</volume>
          <person-group person-group-type="author">
            <name>
              <surname>Altunkan</surname>
              <given-names>S</given-names>
            </name>
            <name>
              <surname>Ilman</surname>
              <given-names>N</given-names>
            </name>
            <name>
              <surname>Kayat&#252;rk</surname>
              <given-names>N</given-names>
            </name>
            <name>
              <surname>Altunkan</surname>
              <given-names>E</given-names>
            </name>
          </person-group>
        </element-citation>
      </ref>
      <ref id="B12">
        <mixed-citation>12. Coleman A,    Freeman P, Steel S, Shennan A. Validation of the Omron 705IT (HEM-759-E) oscillometric    blood pressure monitoring device according to the British Hypertension Society    protocol. Blood Press Monit. 2006; 11 (1): 27-32.</mixed-citation>
        <element-citation publication-type="journal">
          <article-title>Validation of the Omron 705IT (HEM-759-E) oscillometric blood pressure monitoring device according to the British Hypertension Society protocol</article-title>
          <source>Blood Press Monit</source>
          <date>
            <year>2006</year>
          </date>
          <fpage>27</fpage>
          <lpage>32</lpage>
          <issue>1</issue>
          <volume>11</volume>
          <person-group person-group-type="author">
            <name>
              <surname>Coleman</surname>
              <given-names>A</given-names>
            </name>
            <name>
              <surname>Freeman</surname>
              <given-names>P</given-names>
            </name>
            <name>
              <surname>Steel</surname>
              <given-names>S</given-names>
            </name>
            <name>
              <surname>Shennan</surname>
              <given-names>A</given-names>
            </name>
          </person-group>
        </element-citation>
      </ref>
      <ref id="B13">
        <mixed-citation>13. Topouchian    JA, El Assaad MA, Orobinskaia LV, El Feghali RN, Asmar RG. Validation of two    automatic devices for self-measurement of blood pressure according to the International    Protocol of the European Society of Hypertension: the Omron M6 (HEM-7001-E)    and the Omron R7 (HEM 637-IT). Blood Press Monit. 2006; 11 (3): 165-71.</mixed-citation>
        <element-citation publication-type="journal">
          <article-title>Validation of two automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension: the Omron M6 (HEM-7001-E) and the Omron R7 (HEM 637-IT)</article-title>
          <source>Blood Press Monit</source>
          <date>
            <year>2006</year>
          </date>
          <fpage>165</fpage>
          <lpage>71</lpage>
          <issue>3</issue>
          <volume>11</volume>
          <person-group person-group-type="author">
            <name>
              <surname>Topouchian</surname>
              <given-names>JA</given-names>
            </name>
            <name>
              <surname>El Assaad</surname>
              <given-names>MA</given-names>
            </name>
            <name>
              <surname>Orobinskaia</surname>
              <given-names>LV</given-names>
            </name>
            <name>
              <surname>El Feghali</surname>
              <given-names>RN</given-names>
            </name>
            <name>
              <surname>Asmar</surname>
              <given-names>RG</given-names>
            </name>
          </person-group>
        </element-citation>
      </ref>
      <ref id="B14">
        <mixed-citation>14. Plavnik FL,    Zanella MT. Estudo de valida&#231;&#227;o do monitor autom&#225;tico Omron    modelo HEM-608 comparado com o m&#233;todo convencional de medi&#231;&#227;o    de press&#227;o arterial. Arq Bras Cardiol. 2001; 77 (6): 532-6.</mixed-citation>
        <element-citation publication-type="journal">
          <article-title>Estudo de valida&#231;&#227;o do monitor autom&#225;tico Omron modelo HEM-608 comparado com o m&#233;todo convencional de medi&#231;&#227;o de press&#227;o arterial</article-title>
          <source>Arq Bras Cardiol</source>
          <date>
            <year>2001</year>
          </date>
          <fpage>532</fpage>
          <lpage>6</lpage>
          <issue>6</issue>
          <volume>77</volume>
          <person-group person-group-type="author">
            <name>
              <surname>Plavnik</surname>
              <given-names>FL</given-names>
            </name>
            <name>
              <surname>Zanella</surname>
              <given-names>MT</given-names>
            </name>
          </person-group>
        </element-citation>
      </ref>
      <ref id="B15">
        <mixed-citation>15. Furusawa EA,    Ruiz MF, Saito MI, Koch VH. Evaluation of the Omron 705-CP blood pressure measuring    device for use in adolescents and young adults. Arq Bras Cardiol. 2005; 84 (5):    367-70.</mixed-citation>
        <element-citation publication-type="journal">
          <article-title>Evaluation of the Omron 705-CP blood pressure measuring device for use in adolescents and young adults</article-title>
          <source>Arq Bras Cardiol</source>
          <date>
            <year>2005</year>
          </date>
          <fpage>367</fpage>
          <lpage>70</lpage>
          <issue>5</issue>
          <volume>84</volume>
          <person-group person-group-type="author">
            <name>
              <surname>Furusawa</surname>
              <given-names>EA</given-names>
            </name>
            <name>
              <surname>Ruiz</surname>
              <given-names>MF</given-names>
            </name>
            <name>
              <surname>Saito</surname>
              <given-names>MI</given-names>
            </name>
            <name>
              <surname>Koch</surname>
              <given-names>VH</given-names>
            </name>
          </person-group>
        </element-citation>
      </ref>
      <ref id="B16">
        <mixed-citation>16. Bald M, Bohn    W, Feldhoff C, Bonzel KE. Home blood pressure self measurement in children and    adolescents with renal replacement therapy. Klin Padiatr. 2001; 213 (1): 21-5.</mixed-citation>
        <element-citation publication-type="journal">
          <article-title>Home blood pressure self measurement in children and adolescents with renal replacement therapy</article-title>
          <source>Klin Padiatr</source>
          <date>
            <year>2001</year>
          </date>
          <fpage>21</fpage>
          <lpage>5</lpage>
          <issue>1</issue>
          <volume>213</volume>
          <person-group person-group-type="author">
            <name>
              <surname>Bald</surname>
              <given-names>M</given-names>
            </name>
            <name>
              <surname>Bohn</surname>
              <given-names>W</given-names>
            </name>
            <name>
              <surname>Feldhoff</surname>
              <given-names>C</given-names>
            </name>
            <name>
              <surname>Bonzel</surname>
              <given-names>KE</given-names>
            </name>
          </person-group>
        </element-citation>
      </ref>
      <ref id="B17">
        <mixed-citation>17. Silva MAM,    Rivera IR, Ferraz MRMT, Pinheiro AJT, Alves SWS, Moura AA, et al. Prevalence    of cardiovascular risk factors in child and adolescent students in the city    of Macei&#243;. Arq Bras Cardiol. 2005; 84 (5): 387-92.</mixed-citation>
        <element-citation publication-type="journal">
          <article-title>Prevalence of cardiovascular risk factors in child and adolescent students in the city of Macei&#243;</article-title>
          <source>Arq Bras Cardiol</source>
          <date>
            <year>2005</year>
          </date>
          <fpage>387</fpage>
          <lpage>92</lpage>
          <issue>5</issue>
          <volume>84</volume>
          <person-group person-group-type="author">
            <name>
              <surname>Silva</surname>
              <given-names>MAM</given-names>
            </name>
            <name>
              <surname>Rivera</surname>
              <given-names>IR</given-names>
            </name>
            <name>
              <surname>Ferraz</surname>
              <given-names>MRMT</given-names>
            </name>
            <name>
              <surname>Pinheiro</surname>
              <given-names>AJT</given-names>
            </name>
            <name>
              <surname>Alves</surname>
              <given-names>SWS</given-names>
            </name>
            <name>
              <surname>Moura</surname>
              <given-names>AA</given-names>
            </name>
          </person-group>
        </element-citation>
      </ref>
      <ref id="B18">
        <mixed-citation>18. Rezende DF,    Scarpelli RA, de Souza GF, da Costa JO, Scarpelli AM, Scarpelli PA, et al. Prevalence    of systemic hypertension in students aged 7 to 14 years in the municipality    of Barbacena, in the State of Minas Gerais, in 1999. Arq Bras Cardiol. 2003;    81 (4): 381-6.</mixed-citation>
        <element-citation publication-type="journal">
          <article-title>Prevalence of systemic hypertension in students aged 7 to 14 years in the municipality of Barbacena, in the State of Minas Gerais, in 1999</article-title>
          <source>Arq Bras Cardiol</source>
          <date>
            <year>2003</year>
          </date>
          <fpage>381</fpage>
          <lpage>6</lpage>
          <issue>4</issue>
          <volume>81</volume>
          <person-group person-group-type="author">
            <name>
              <surname>Rezende</surname>
              <given-names>DF</given-names>
            </name>
            <name>
              <surname>Scarpelli</surname>
              <given-names>RA</given-names>
            </name>
            <name>
              <surname>de Souza</surname>
              <given-names>GF</given-names>
            </name>
            <name>
              <surname>da Costa</surname>
              <given-names>JO</given-names>
            </name>
            <name>
              <surname>Scarpelli</surname>
              <given-names>AM</given-names>
            </name>
            <name>
              <surname>Scarpelli</surname>
              <given-names>PA</given-names>
            </name>
          </person-group>
        </element-citation>
      </ref>
      <ref id="B19">
        <mixed-citation>19. Costa JSD,    Barcellos FC, Sclowitz ML, Sclowitz IKT, Castanheira M, Olinto MTA, et al. Preval&#234;ncia    de hipertens&#227;o arterial em adultos e fatores associados: um estudo de    base populacional urbana em Pelotas, Rio Grande do Sul, Brasil. Arq Bras Cardiol.    2007; 88 (1): 59-65.</mixed-citation>
        <element-citation publication-type="journal">
          <article-title>Preval&#234;ncia de hipertens&#227;o arterial em adultos e fatores associados: um estudo de base populacional urbana em Pelotas, Rio Grande do Sul, Brasil</article-title>
          <source>Arq Bras Cardiol</source>
          <date>
            <year>2007</year>
          </date>
          <fpage>59</fpage>
          <lpage>65</lpage>
          <issue>1</issue>
          <volume>88</volume>
          <person-group person-group-type="author">
            <name>
              <surname>Costa</surname>
              <given-names>JSD</given-names>
            </name>
            <name>
              <surname>Barcellos</surname>
              <given-names>FC</given-names>
            </name>
            <name>
              <surname>Sclowitz</surname>
              <given-names>ML</given-names>
            </name>
            <name>
              <surname>Sclowitz</surname>
              <given-names>IKT</given-names>
            </name>
            <name>
              <surname>Castanheira</surname>
              <given-names>M</given-names>
            </name>
            <name>
              <surname>Olinto</surname>
              <given-names>MTA</given-names>
            </name>
          </person-group>
        </element-citation>
      </ref>
      <ref id="B20">
        <mixed-citation>20. Souza ARA,    Costa A, Nakamura D, Mocheti LM, Stevanato Filho PR, Ovando LA. Um estudo sobre    hipertens&#227;o arterial sist&#234;mica na cidade de Campo Grande, MS. Arq    Bras Cardiol. 2007; 88 (4): 441-6.</mixed-citation>
        <element-citation publication-type="journal">
          <article-title>Um estudo sobre hipertens&#227;o arterial sist&#234;mica na cidade de Campo Grande, MS</article-title>
          <source>Arq Bras Cardiol</source>
          <date>
            <year>2007</year>
          </date>
          <fpage>441</fpage>
          <lpage>6</lpage>
          <issue>4</issue>
          <volume>88</volume>
          <person-group person-group-type="author">
            <name>
              <surname>Souza</surname>
              <given-names>ARA</given-names>
            </name>
            <name>
              <surname>Costa</surname>
              <given-names>A</given-names>
            </name>
            <name>
              <surname>Nakamura</surname>
              <given-names>D</given-names>
            </name>
            <name>
              <surname>Mocheti</surname>
              <given-names>LM</given-names>
            </name>
            <name>
              <surname>Stevanato Filho</surname>
              <given-names>PR</given-names>
            </name>
            <name>
              <surname>Ovando</surname>
              <given-names>LA</given-names>
            </name>
          </person-group>
        </element-citation>
      </ref>
      <ref id="B21">
        <mixed-citation>21. Gus I, Fischmann    A, Medina C. Prevalence of risk factors for coronary artery disease in the Brazilian    State of Rio Grande do Sul. Arq Bras Cardiol. 2002; 78 (5): 478-90.</mixed-citation>
        <element-citation publication-type="journal">
          <article-title>Prevalence of risk factors for coronary artery disease in the Brazilian State of Rio Grande do Sul</article-title>
          <source>Arq Bras Cardiol</source>
          <date>
            <year>2002</year>
          </date>
          <fpage>478</fpage>
          <lpage>90</lpage>
          <issue>5</issue>
          <volume>78</volume>
          <person-group person-group-type="author">
            <name>
              <surname>Gus</surname>
              <given-names>I</given-names>
            </name>
            <name>
              <surname>Fischmann</surname>
              <given-names>A</given-names>
            </name>
            <name>
              <surname>Medina</surname>
              <given-names>C</given-names>
            </name>
          </person-group>
        </element-citation>
      </ref>
      <ref id="B22">
        <mixed-citation>22. Feij&#227;o    AM, Gadelha FV, Bezerra AA, de Oliveira AM, Silva Mdo S, Lima JW. Prevalence    of excessive weight and hypertension in a low-income urban population. Arq Bras    Cardiol. 2005; 84 (1): 29-33.</mixed-citation>
        <element-citation publication-type="journal">
          <article-title>Prevalence of excessive weight and hypertension in a low-income urban population</article-title>
          <source>Arq Bras Cardiol</source>
          <date>
            <year>2005</year>
          </date>
          <fpage>29</fpage>
          <lpage>33</lpage>
          <issue>1</issue>
          <volume>84</volume>
          <person-group person-group-type="author">
            <name>
              <surname>Feij&#227;o</surname>
              <given-names>AM</given-names>
            </name>
            <name>
              <surname>Gadelha</surname>
              <given-names>FV</given-names>
            </name>
            <name>
              <surname>Bezerra</surname>
              <given-names>AA</given-names>
            </name>
            <name>
              <surname>de Oliveira</surname>
              <given-names>AM</given-names>
            </name>
            <name>
              <surname>Silva Mdo</surname>
              <given-names>S</given-names>
            </name>
            <name>
              <surname>Lima</surname>
              <given-names>JW</given-names>
            </name>
          </person-group>
        </element-citation>
      </ref>
      <ref id="B23">
        <mixed-citation>23. Park MK, Menard    SW, Yuan C. Comparison of auscultatory and oscillometric blood pressures. Arch    Pediatr Adolesc Med. 2001; 155 (1): 50-3.</mixed-citation>
        <element-citation publication-type="journal">
          <article-title>Comparison of auscultatory and oscillometric blood pressures</article-title>
          <source>Arch Pediatr Adolesc Med</source>
          <date>
            <year>2001</year>
          </date>
          <fpage>50</fpage>
          <lpage>3</lpage>
          <issue>1</issue>
          <volume>155</volume>
          <person-group person-group-type="author">
            <name>
              <surname>Park</surname>
              <given-names>MK</given-names>
            </name>
            <name>
              <surname>Menard</surname>
              <given-names>SW</given-names>
            </name>
            <name>
              <surname>Yuan</surname>
              <given-names>C</given-names>
            </name>
          </person-group>
        </element-citation>
      </ref>
    </ref-list>
  </back>
  <sub-article article-type="translation" id="TRen" xml:lang="en">
    <front-stub>
      <article-categories>
        <subj-group subj-group-type="heading">
          <subject>Original Articles</subject>
        </subj-group>
      </article-categories>
      <title-group>
        <article-title xml:lang="en">Validation of the omron HEM 742 blood pressure monitoring device in adolescents</article-title>
      </title-group>
      <abstract xml:lang="en">
        <p>BACKGROUND: Accurate blood pressure measurement is fundamental for scientific investigation or clinical decision-making. In this sense, it is important to verify the values provided by electronic devices. OBJECTIVE: To validate the Omron HEM 742 blood pressure monitoring device in adolescents according to criteria suggested by the British Hypertension Society. METHODS: A total of 150 adolescents aged between 10 and 16 years participated in the study. The automated Omron HEM 742 monitor was connected in Y to the mercury column auscultatory device, then three simultaneous measurements were taken, and the differences between the readings of the two devices were calculated. The intraclass correlation coefficient and Bland-Altman plot (agreement) were used to verify the relationship between both devices. Specificity and sensitivity of the device were determined by using the ROC curve. RESULTS: The comparison between the measurements showed an equal to or lower than 5mmHg difference in 67.3% of the systolic values, and 69.3% of the diastolic values; an equal to or lower than 10mmHg difference occurred in 87.3% and 90.6% of the systolic and diastolic values, respectively; an equal to or lower than 15mmHg difference was found in 96.6% of the systolic values and 97.3% of the diastolic values. These findings are consistent with a grade A according to the British Hypertension Society protocol. A marked consonance was observed between the values obtained by the automated monitor and this device was proven to be capable of identifying the presence or absence of high blood pressure. CONCLUSION: The Omron HEM 742 monitor was proved valid for blood pressure measurement in adolescents according to the criteria suggested by the British Hypertension Society.</p>
      </abstract>
      <kwd-group xml:lang="en">
        <kwd>Blood pressure</kwd>
        <kwd>blood pressure monitors</kwd>
        <kwd>adolescent</kwd>
      </kwd-group>
    </front-stub>
    <body>
      <p>
        <bold>ORIGINAL ARTICLE</bold>
      </p>
      <p>
        <fn id="top-body2">
          <label>
            <bold>Validation of the omron HEM 742 blood pressure monitoring device in adolescents</bold>
          </label>
          <p>
            <bold>Diego Giulliano Destro Christofaro<sup>I</sup>; R&#244;mulo Ara&#250;jo Fernandes<sup>II</sup>; Aline Mendes Gerage<sup>I</sup>; Marcelo Jos&#233; Alves<sup>II</sup>; Marcos Doederlein Polito<sup>I</sup>; Arli Ramos de Oliveira<sup>I</sup></bold>
          </p>
        </fn>
      </p>
      <p><sup>I</sup>Universidade Estadual de Londrina, Londrina, PR</p>
      <p><sup>II</sup>Universidade Estadual Paulista, S&#227;o Paulo, SP - Brazil</p>
      <p>Mailing address</p>
      <p>
        <bold>SUMMARY</bold>
      </p>
      <p><bold>BACKGROUND: </bold> Accurate blood pressure measurement is fundamental for scientific investigation or clinical decision-making. In this sense, it is important to verify the values provided by electronic devices.</p>
      <p><bold>OBJECTIVE: </bold> To validate the Omron HEM 742 blood pressure monitoring device in adolescents according to criteria suggested by the British Hypertension Society.</p>
      <p><bold>METHODS: </bold> A total of 150 adolescents aged between 10 and 16 years participated in the study. The automated Omron HEM 742 monitor was connected in Y to the mercury column auscultatory device, then three simultaneous measurements were taken, and the differences between the readings of the two devices were calculated. The intraclass correlation coefficient and Bland-Altman plot (agreement) were used to verify the relationship between both devices. Specificity and sensitivity of the device were determined by using the ROC curve.</p>
      <p><bold>RESULTS: </bold> The comparison between the measurements showed an equal to or lower than 5mmHg difference in 67.3% of the systolic values, and 69.3% of the diastolic values; an equal to or lower than 10mmHg difference occurred in 87.3% and 90.6% of the systolic and diastolic values, respectively; an equal to or lower than 15mmHg difference was found in 96.6% of the systolic values and 97.3% of the diastolic values. These findings are consistent with a grade A according to the British Hypertension Society protocol. A marked consonance was observed between the values obtained by the automated monitor and this device was proven to be capable of identifying the presence or absence of high blood pressure.</p>
      <p><bold>CONCLUSION: </bold> The Omron HEM 742 monitor was proved valid for blood pressure measurement in adolescents according to the criteria suggested by the British Hypertension Society.</p>
      <p><bold>Key words:</bold> Blood pressure; blood pressure monitors/utilization; adolescent.</p>
      <p>
        <bold>Introduction</bold>
      </p>
      <p>The prevalence of high blood pressure (BP) has increased steeply among adult men and women from various countries<sup>1</sup>, and is a cause of concern for health care professionals, since this condition is considered one of the major risk factors for the development of cardiovascular diseases<sup>2,3</sup>. However, national data show that children and adolescents may present significantly increased BP levels<sup>4,5</sup>, which is also disturbing, considering that the risk conditions during childhood and adolescence tend to be expressed in adulthood<sup>6</sup>.</p>
      <p>Accurate and reliable BP measurements, whether in the clinical practice or for experimental purposes, are indispensable to help in the diagnosis or interpretation of findings. In this sense, the literature points out several forms of BP measurement<sup>7</sup>. Among these methods, the auscultatory technique is the one that has been most commonly used, whether by means of aneroid or mercury sphygmomanometers. However, the accuracy of this method may be affected by factors relative to the observer.</p>
      <p>Thus, an alternative for BP measurement is the utilization of electronic devices. To date, a growing number of such devices has been found in the market, considering that they are affordable and easy to use. However, it is important to evaluate these devices according to validation rules required by international entities such as the British Hypertension Society (BHS)<sup>8</sup>. Data available in the literature regarding the validity of automated devices were mainly obtained from samples of adult individuals. Trials involving children and adolescents are fewer and thus suggest further investigation.</p>
      <p>Therefore, the objective of the present study was to validate the Omron HEM 742 oscillometric automated BP monitoring device in children and adolescents according to the protocol proposed by the BHS, and to evaluate its efficacy in the classification of individuals within this age range in relation to resting BP values.</p>
      <p>
        <bold>Methods</bold>
      </p>
      <p>
        <bold>Sample</bold>
      </p>
      <p>Data were collected in an institution of the state education system. The school unit was selected by convenience for meeting all the requirements pre-established by the researchers</p>
      <p><bold>a)</bold> proximity to the Higher Education Institution;</p>
      <p><bold>b)</bold> facilities adequate for measurement taking;</p>
      <p><bold>c)</bold> authorization from the principal in charge of the school unit;</p>
      <p><bold>d)</bold> meeting the age range previously defined by the researchers).</p>
      <p>All subjects who, according to information provided by their guardians, had arrhythmias or atrial fibrillation, had ingested caffeine 30 min prior to the assessment or were using chronotropic or inotropic agents were excluded from the sample analyzed. All sample subjects informed not to be tobacco smokers and they did not perform physical activities on the day their blood pressure was measured. Thus, the sample was comprised of 150 adolescents (77 males and 73 females).</p>
      <p>All guardians signed a written informed consent agreeing with the participation of the subjects in the study, according to instructions contained in Resolution 196/96 of the National Health Council for studies in humans. The study was approved by the Ethics Committee of the Higher Education Institution.</p>
      <p>
        <bold>Anthropometric measurements</bold>
      </p>
      <p>The chronologic age of the adolescents was determined in the hundredth form using the date of birth and day of assessment. Body mass and height were measured to characterize the sample. For this purpose, a portable digital scale (graduation: 100g and maximum capacity: 150Kg) and a metal stadiometer fixed to the wall (accuracy: 0.1cm and maximum height: two meters) were used. All measurements were taken with the individuals barefoot and dressing light clothes, according to the protocol suggested by Gordon et al<sup>9</sup>. Based on this information, the body mass index (BMI) was calculated, and the values were expressed as kilograms per square meter (kg/m<sup>2</sup>).</p>
      <p>
        <bold>Blood pressure measurements</bold>
      </p>
      <p>For BP measurement of the adolescents, two types of cuffs were used according to the arm circumference (6mm x 12mm [child size], and 9mm x 18mm for adolescents 14-16 years old and those younger but who had a greater body size]), following the recommendations of the American Heart Association<sup>7</sup>. All blood pressure measurements were taken in the right arm of each individual. The device to be tested was the Omron HEM 742, which consists of an electronic and digital device for arm BP measurement with automated air inflation and deflation. This device uses the oscillometric measurement method, with pressure range from 0-280mmHg.</p>
      <p>The oscillometric device was connected in "Y" to the mercury sphygmomanometer. The inflation mechanism of the device was activated to perform three consecutive measurements with a two-minute interval between each. At the end of each measurement, an observer recorded systolic (SBP) and diastolic blood pressure (DBP) readings identified in the mercury column, blind to the readings identified in the automated device and recorded by another observer.</p>
      <p>Before being assessed, the individuals remained sitting at rest for five minutes with their trunk leaning against a chair and the arm relaxed. The mercury column and the automated device were duly calibrated before the measurements were taken. The time spent for the full assessment of each sample subject (anthropometric and blood pressure assessment) was approximately 10 minutes.</p>
      <p>A total of 450 simultaneous assessments were made with both BP measurement devices. Classification of the variation of BP values provided by the device was made considering the differences of the reading records between the mercury manometer and the automated device according to the procedures described by the BHS. The BHS protocol suggests that devices should achieve at least grade B for SBP and DBP to be deemed valid. The respective grades and their values are described in <xref ref-type="fig" rid="box1-body2">Box 1</xref>.</p>
      <fig id="box1-body2">
        <graphic xlink:href="1678-4170-abc-92-01-0010-00003-gf01-en.jpg"/></fig>
      <p>
        <bold>Statistical analysis</bold>
      </p>
      <p>Initially, the Kolmogorov-Smirnov test was used to verify data distribution. Then, descriptive analysis was performed using mean and standard deviation (SD) values, since all the variables analyzed fit the Gaussian distribution model. Analysis of variance for repeated measures (with Greenhouse-Geisser correction when the sphericity assumption was violated) and the Student's t test for paired samples were used for comparisons between the mean BP values found by the two devices. In order to verify the relationship between the BP values provided by the automated device and those provided by the mercury device, the intraclass correlation coefficient (ICC) was used. Agreement measurements were tested using the Bland-Altman plot (numeric values), and the parameters provided by the ROC curve (sensitivity, specificity and area under the curve [AUC]) indicated the power of diagnosis of high blood pressure values of the oscillometric device. Significance values lower than 5% were considered statistically significant. The statistical software programs used were MedCalc and SPSS version 13.0.</p>
      <p>
        <bold>Results</bold>
      </p>
      <p>The overall sample characteristics grouped according to gender are shown in <xref ref-type="table" rid="tab1-body2">Table 1</xref>. No statistically significant differences were found between the variables analyzed.</p>
      <table-wrap id="tab1-body2">
        <graphic xlink:href="1678-4170-abc-92-01-0010-00003-gt01-en.jpg"/></table-wrap>
      <p>The results presented in <xref ref-type="table" rid="tab2-body2">Table 2</xref> show the differences between the measurements obtained by the automated and the mercury devices in percentage values. According to the results found, the device tested received grade A both for SBP and DBP, as per the criteria established by the BHS.</p>
      <table-wrap id="tab2-body2">
        <graphic xlink:href="1678-4170-abc-92-01-0010-00003-gt02-en.jpg"/></table-wrap>
      <p>The comparisons and the ICC for SBP and DBP values are shown in <xref ref-type="table" rid="en_03t3-body2">Table 3</xref>. Significant differences for SBP values were found between the devices used and in all three measurements. However, for DBP, difference was found only in the first measurement. For both devices, no statistically significant differences were found for SBP and DBP values between the three measurements taken. As regards gender, the correlation values for SBP observed between the devices ranged from moderate to high (ICC = 0.74 and ICC = 0.88). For DBP, in turn, moderate values were observed (ICC = 0.50 and ICC = 0.79). In both circumstances, no differences were found between genders that could compromise the use of the device.</p>
      <p content-type="asset">
        <table-wrap id="en_03t3-body2">
          <graphic xlink:href="1678-4170-abc-92-01-0010-00003-gt03-en.jpg"/></table-wrap>
      </p>
      <p>The Bland-Altman plot (<xref ref-type="fig" rid="fig-body2">Figure 1</xref>) was used as an agreement indicator for BP values provided by both devices. For SBP values, only 2.2% of the 450 measurements were observed to be out of the 95% confidence interval set for the mean of the differences observed. For DBP values, in turn, only 2.4% of the measurements were out of this confidence interval.</p>
      <fig id="fig-body2">
        <graphic xlink:href="1678-4170-abc-92-01-0010-00003-gf02-en.jpg"/></fig>
      <p>After classification of the individuals according to their hemodynamic behavior at rest (normal blood pressure / high BP), the efficacy of the automated device in diagnosing the presence of high BP was tested in the 450 measurements (<xref ref-type="table" rid="tab4-body2">Table 4</xref>).</p>
      <table-wrap id="tab4-body2">
        <label>Table 4</label>
        <graphic xlink:href="1678-4170-abc-92-01-0010-00003-gt04-en.jpg"/></table-wrap>
      <p>
        <bold>Discussion</bold>
      </p>
      <p>The purpose of this study was to validate the Omrom HEM 742 oscillometric automated device for BP measurement in children and adolescents according to the protocol proposed by the BHS, and to evaluate its efficacy in the classification of individuals of this age range in relation to BP values at rest, considering that this device has been validated only in adult individuals<sup>10</sup> and that the majority of the studies involving the validation of other automated devices for BP measurement was also performed with adult populations<sup>11-14</sup>. Additionally, with the increasing occurrence of high BP values at rest that has been observed among children and adolescents<sup>4,5</sup>, studies with younger populations are also required so as to create strategies to facilitate BP measurement and its interpretation in this population, both in the clinical practice and in research.</p>
      <p>Thereby, we should point out that automated inflation and deflation devices are light, portable and easy to use and interpret, thus reducing the occurrence of reading errors by the observer, which is relatively often seen with the auscultatory method. Therefore, the use of automated devices seems to be an interesting strategy for BP measurement, especially when related to a better home blood pressure control by hypertensive populations.</p>
      <p>For a device to be validated for use in special populations, the BHS<sup>8</sup> suggests that it should be previously validated in adults, as is the case of the Omron HEM 742 device validated by Coleman et al<sup>10</sup>, and also that different BP values ranging from 100mmHg to 240mmHg for SBP and from 60mmHg to 120mmHg for DBP should be evaluated. However, this was a study limitation, since our values ranged from 98mmHg to 135mmHg, and from 53mmHg to 86mmHg for SBP and DBP, respectively. Nonetheless, this variation verified in the present investigation is similar to that found in other studies of validation of devices for BP measurement in young populations<sup>15,16</sup>, in which values higher than 140mmHg for SBP and 90mmHg for DBP were not found.</p>
      <p>This is mainly due to the fact that prevalence rates of hypertension are lower in this age range<sup>17,18</sup> in comparison to the adult population<sup>18-22</sup>. Also, we should point out that the diagnosis of hypertension in the young population is based on percentiles considering gender, age and height, unlike the diagnosis in adults for which there are pre-established SBP and DBP values.</p>
      <p>In the present study, in all three measurements taken in the sample, the SBP values provided by the device tested were significantly higher when compared with the values obtained by the mercury column device. These higher values obtained with the oscillometric device may have resulted from lower measurements obtained by the observer using the auscultatory method and/or they may have really been higher values obtained with the test device<sup>23</sup>. However, a definitive explanation for this behavior has not been fully defined in the literature. On the other hand, in relation to DBP, a significant difference was only found in the first measurement. Despite these differences, the SBP and DBP values provided by the Omron HEM 742 and by the mercury device were not significantly different between the three measurements when compared with each other, thus suggesting good data reproducibility in both devices.</p>
      <p>The ICC was significant and showed moderate/high values in all measurements, both for SBP and DBP, thus indicating a good relationship between the values observed for both devices. That is, the higher the value obtained with a device, the higher the value found by the other device for the same individual. However, this statistical procedure is only a relationship measurement and does not indicate the agreement between these two devices and, consequently, does not detect possible extreme variations between them. This limitation is overcome by the use of the Bland-Altman plot which does indicate the agreement between numeric data. In the present study, low disagreement was observed between the BP values found by both devices. This fact was confirmed when we verified that the criteria proposed by the BHS<sup>8</sup> had been met and, also, that the highest grade had been achieved.</p>
      <p>These positive indicators shown by the ICC and by the Bland-Altman plot are important information on the reliability of the use of the device tested. However, these are numeric values and, therefore, they are not able to indicate the efficacy of the device in the clinical practice, more specifically in the detection of high BP values. To this end, the use of the ROC curve is required. Thus, the parameters provided by the ROC curve showed that the Omron HEM 742 device was highly capable of detecting both the presence (sensitivity) and the absence (specificity) of high BP levels in the sample studied.</p>
      <p>
        <bold>Conclusion</bold>
      </p>
      <p>The Omrom HEM 742 device has good reliability for use in the assessment of adolescents, since it was validated according to international recommendations of the BHS<sup>8</sup>, and achieved grade A for SBP and DBP analyses, in addition to having presented good sensitivity and specificity in the diagnosis of blood pressure levels in individuals within this age range.</p>
      <p>
        <bold>Potential Conflict of Interest</bold>
      </p>
      <p>No potential conflict of interest relevant to this article was reported.</p>
      <p>
        <bold>Sources of Funding</bold>
      </p>
      <p>There were no external funding sources for this study.</p>
      <p>
        <bold>Study Association</bold>
      </p>
      <p>This article is part of the thesis of master submitted by Diego Giulliano Destro Christofaro, from Universidade Estadual de Londrina.</p>
      <p>
        <bold>References</bold>
      </p>
    </body>
  </sub-article>
  <sub-article article-type="translation" id="TRes" xml:lang="es">
    <front-stub>
      <article-categories>
        <subj-group subj-group-type="heading">
          <subject>Art&#237;culos Originales</subject>
        </subj-group>
      </article-categories>
      <title-group>
        <article-title xml:lang="es">Validaci&#243;n del monitor de medida de presi&#243;n arterial Omron HEM 742 en adolescentes</article-title>
      </title-group>
      <abstract xml:lang="es">
        <p>FUNDAMENTO: La medida precisa de la presi&#243;n arterial es fundamental para la investigaci&#243;n cient&#237;fica o decisi&#243;n cl&#237;nica. En este sentido, es importante que se verifique valores provenientes de equipos electr&#243;nicos. OBJETIVO: Validar el monitor Omron HEM 742 de medida de presi&#243;n arterial en adolescentes, de acuerdo con los criterios sugeridos por British Hypertension Society. M&#201;TODOS: Participaron del estudio 150 adolescentes con edades entre 10 y 16 a&#241;os. Se conect&#243; el monitor autom&#225;tico Omron HEM 742 en Y con equipo de auscultaci&#243;n de columna de mercurio, y se realizaron tres evaluaciones simult&#225;neas, calcul&#225;ndose las diferencias entre los dos equipos. Para verificarse la relaci&#243;n entre ambos, se utiliz&#243; el coeficiente de correlaci&#243;n intraclase y el m&#233;todo de Bland-Altman (concordancia). Se determin&#243; la especificidad y la sensibilidad del aparato por medio de la curva ROC. RESULTADOS: La comparaci&#243;n entre las medidas evidenci&#243; una diferencia menor o igual a 5 mmHg en el 67,3% de los valores sist&#243;licos y el 69,3% de los valores diast&#243;licos; una diferencia &lt; a 10 mmHg ocurri&#243; en el 87,3% y el 90,6% de los valores sist&#243;licos y diast&#243;licos, respectivamente; y una diferencia &lt; a 15 mmHg en el 96,6% de los valores sist&#243;licos y el 97,3% de los diast&#243;licos. Esos resultados indican grado A seg&#250;n el protocolo de British Hypertension Society. Se observ&#243; a&#250;n elevada concordancia en los valores obtenidos por medio del monitor autom&#225;tico, y se verific&#243; que este equipo es capaz de identificar la presencia o la ausencia de la presi&#243;n arterial elevada. CONCLUSI&#211;N: El monitor Omron HEM 742 se mostr&#243; v&#225;lido para medidas de presi&#243;n arterial en adolescentes, de acuerdo con los criterios sugeridos por British Hypertension Society.</p>
      </abstract>
      <kwd-group xml:lang="es">
        <kwd>Presi&#243;n arterial</kwd>
        <kwd>monitores de presi&#243;n arterial</kwd>
        <kwd>adolescente</kwd>
      </kwd-group>
    </front-stub>
    <body>
      <p>
        <bold>ART&#205;CULO ORIGINAL</bold>
      </p>
      <p>
        <fn id="top-body3">
          <label>
            <bold>Validaci&#243;n del monitor de medida de presi&#243;n arterial Omron HEM 742 en adolescentes </bold>
          </label>
          <p>
            <bold>Diego Giulliano Destro Christofaro<sup>I</sup>; R&#244;mulo Ara&#250;jo Fernandes<sup>II</sup>; Aline Mendes Gerage<sup>I</sup>; Marcelo Jos&#233; Alves<sup>II</sup>; Marcos Doederlein Polito<sup>I</sup>; Arli Ramos de Oliveira<sup>I</sup></bold>
          </p>
        </fn>
      </p>
      <p><sup>I</sup>Universidade Estadual de Londrina, Londrina, PR</p>
      <p><sup>II</sup>Universidade Estadual Paulista, S&#227;o Paulo, SP - Brasil</p>
      <p>Correspondencia</p>
      <p>
        <bold>RESUMEN</bold>
      </p>
      <p><bold>FUNDAMENTO: </bold> La medida precisa de la presi&#243;n arterial es fundamental para la investigaci&#243;n cient&#237;fica o decisi&#243;n cl&#237;nica. En este sentido, es importante que se verifique valores provenientes de equipos electr&#243;nicos.</p>
      <p><bold>OBJETIVO:</bold> Validar el monitor Omron HEM 742 de medida de presi&#243;n arterial en adolescentes, de acuerdo con los criterios sugeridos por <italic>British Hypertension Society</italic>.</p>
      <p><bold>M&#201;TODOS:</bold> Participaron del estudio 150 adolescentes con edades entre 10 y 16 a&#241;os. Se conect&#243; el monitor autom&#225;tico Omron HEM 742 en Y con equipo de auscultaci&#243;n de columna de mercurio, y se realizaron tres evaluaciones simult&#225;neas, calcul&#225;ndose las diferencias entre los dos equipos. Para verificarse la relaci&#243;n entre ambos, se utiliz&#243; el coeficiente de correlaci&#243;n intraclase y el m&#233;todo de Bland-Altman (concordancia). Se determin&#243; la especificidad y la sensibilidad del aparato por medio de la curva ROC.</p>
      <p><bold>RESULTADOS:</bold> La comparaci&#243;n entre las medidas evidenci&#243; una diferencia menor o igual a 5 mmHg en el 67,3% de los valores sist&#243;licos y el 69,3% de los valores diast&#243;licos; una diferencia <underline>&lt;</underline> a 10 mmHg ocurri&#243; en el 87,3% y el 90,6% de los valores sist&#243;licos y diast&#243;licos, respectivamente; y una diferencia <underline>&lt;</underline> a 15 mmHg en el 96,6% de los valores sist&#243;licos y el 97,3% de los diast&#243;licos. Esos resultados indican grado A seg&#250;n el protocolo de <italic>British Hypertension Society</italic>. Se observ&#243; a&#250;n elevada concordancia en los valores obtenidos por medio del monitor autom&#225;tico, y se verific&#243; que este equipo es capaz de identificar la presencia o la ausencia de la presi&#243;n arterial elevada.</p>
      <p><bold>CONCLUSI&#211;N:</bold> El monitor Omron HEM 742 se mostr&#243; v&#225;lido para medidas de presi&#243;n arterial en adolescentes, de acuerdo con los criterios sugeridos por <italic>British Hypertension Society</italic>. </p>
      <p><bold>Palabras clave: </bold> Presi&#243;n arterial, monitores de presi&#243;n arterial/utilizaci&#243;n, adolescente.</p>
      <p>
        <bold>Introducci&#243;n</bold>
      </p>
      <p>La prevalencia de la presi&#243;n arterial (PA) elevada ha crecido acentuadamente en varones y mujeres adultos de varios pa&#237;ses<sup>1</sup>, y se convirti&#243; en motivo de preocupaci&#243;n para profesionales de salud, una vez que esa funci&#243;n se muestra como uno de los principales factores de riesgo para el desarrollo de enfermedades cardiovasculares<sup>2,3</sup>. Sin embargo, datos nacionales muestran que ni&#241;os y adolescentes pueden presentar valores significativamente elevados de PA<sup>4,5</sup>, lo que tambi&#233;n es preocupante si se toma en consideraci&#243;n que las condiciones de riesgo durante la infancia y la adolescencia tienden a expresarse en la vida adulta<sup>6</sup>.</p>
      <p>Medidas precisas y fidedignas de PA, ya sea en la pr&#225;ctica cl&#237;nica ya sea de forma experimental, se convierten en un procedimiento imprescindible para auxiliar en el diagn&#243;stico o en la interpretaci&#243;n de datos. En este sentido, la literatura apunta diversas formas de mensuraci&#243;n de la PA<sup>7</sup>. Entre esos m&#233;todos, &#233;l que se utiliz&#243; de m&#225;s com&#250;n es el m&#233;todo de auscultaci&#243;n, tanto por esfigmoman&#243;metros aneroides como de columna de mercurio. No obstante, la exactitud de este m&#233;todo puede ser comprometida por factores relativos al observador.</p>
      <p>As&#237;, una alternativa para la aferici&#243;n de la PA es la utilizaci&#243;n de aparatos electr&#243;nicos. Actualmente, se verifica un n&#250;mero creciente de esos aparatos en el mercado, tomando en consideraci&#243;n que, adem&#225;s de poseer precios accesibles, son de f&#225;cil manipulaci&#243;n. Sin embargo, es importante que se eval&#250;e esos equipos de acuerdo con las normas de validaci&#243;n exigidas por entidades internacionales, como la de <italic>British Hypertension Society</italic> (BHS)<sup>8</sup>. Los datos disponibles en la literatura en cuanto a la validad de aparatos autom&#225;ticos, utilizaron principalmente una muestra de individuos adultos. Los experimentos utilizando ni&#241;os y adolescentes son en menor n&#250;mero y, por ello, sugieren mejores investigaciones.</p>
      <p>El presente estudio, por lo tanto, tuvo como objetivo validar el aparato de oscilaci&#243;n autom&#225;tica Omron HEM 742 de medida de PA en ni&#241;os y adolescentes, de acuerdo con el protocolo propuesto por la BHS, adem&#225;s de evaluar su eficiencia en la clasificaci&#243;n de los individuos de esta franja etaria, con relaci&#243;n a los valores de PA de reposo.</p>
      <p>
        <bold>M&#233;todos</bold>
      </p>
      <p>
        <bold>Muestra</bold>
      </p>
      <p>Se efectuaron las recolecciones en una instituci&#243;n de la red estatal de ense&#241;anza. Se seleccion&#243; la unidad escolar, por conveniencia, por rellenar todos los requisitos preestablecidos por los investigadores:</p>
      <p>a) proximidad a la instituci&#243;n de ense&#241;anza superior;</p>
      <p>b) instalaciones adecuadas para las afericiones;</p>
      <p>c) autorizaci&#243;n del director responsable por la unidad escolar;</p>
      <p>d) atender a la franja etaria previamente definida por los investigadores.</p>
      <p>De la muestra analizada, se excluyeron a todos los individuos que, por declaraci&#243;n del responsable, pose&#237;an arritmias o fibrilaci&#243;n atrial, ingirieron cafe&#237;na 30 minutos antes de la evaluaci&#243;n o utilizaban f&#225;rmacos de acci&#243;n crono o inotr&#243;pica. Todos los individuos de la muestra informaron no utilizar tabaco ni realizar actividades f&#237;sicas al d&#237;a de la aferici&#243;n de la presi&#243;n arterial. De este modo, 150 adolescentes compusieron la muestra (77 del sexo masculino y 73 del femenino).</p>
      <p>Todos los responsables por los individuos firmaron un formulario de consentimiento informado concordando en participar del estudio, de acuerdo con las instrucciones contenidas en la Resoluci&#243;n n&#186; 196/96 del Consejo Nacional de Salud, para estudios con seres humanos. El estudio fue aprobado por el Comit&#233; de &#201;tica de la Instituci&#243;n de Ense&#241;anza Superior.</p>
      <p>
        <bold>Medidas antropom&#233;tricas</bold>
      </p>
      <p>Se determin&#243; la edad cronol&#243;gica de los adolescentes en forma centesimal, utiliz&#225;ndose la fecha de nacimiento y el d&#237;a de la evaluaci&#243;n. Se realizaron medidas de masa corporal y de estatura para la caracterizaci&#243;n de la muestra. Para ello, se utiliz&#243; una balanza port&#225;til digital (gradaci&#243;n: 100 g y capacidad m&#225;xima: 150 kg) y un estadi&#243;metro de metal fijado a la pared (precisi&#243;n: 0,1 cm y extensi&#243;n m&#225;xima: 2 metros), y todas las medidas se efectuaron con los individuos descalzos y llevando ropas ligeras, de acuerdo con el protocolo sugerido por Gordon et al<sup>9</sup>. Tomando en cuenta estas informaciones, se calcul&#243; el &#237;ndice de masa corporal (IMC), expres&#225;ndose los valores en quilogramos por metro cuadrado (kg/m<sup>2</sup>).</p>
      <p>
        <bold>Medidas de presi&#243;n arterial</bold>
      </p>
      <p>Para la mensuraci&#243;n de los valores de PA entre los adolescentes, se utilizaron dos tipos de manguitos de acuerdo con la circunferencia del brazo (6 mm x 12 mm [talla infantil] y 9 mm x 18 mm [para adolescentes de 14-16 a&#241;os y para aquellos con edad inferior que tuvieran una estructura f&#237;sica mayor]), de acuerdo con las recomendaciones de la Asociaci&#243;n Americana del Coraz&#243;n<sup>7</sup>. Se realizaron todas las evaluaciones de la medida de la presi&#243;n arterial en el brazo derecho de cada individuo. El equipo probado fue el Omron HEM 742, que consiste en un aparato electr&#243;nico y digital de medida de PA de brazo, con inflaci&#243;n y deflaci&#243;n autom&#225;tica del aire. El m&#233;todo de medida de este aparato es el oscilom&#233;trico, con variaci&#243;n de la presi&#243;n de 0-280 mmHg.</p>
      <p>Se conect&#243; el aparato oscilom&#233;trico en Y al esfigmoman&#243;metro de columna de mercurio. Se activ&#243; el mecanismo de inflaci&#243;n del aparato para realizaci&#243;n de tres medidas consecutivas con intervalo de 2 minutos entre ambas. Al final de cada medida, un observador registraba los valores de la presi&#243;n arterial sist&#243;lica (PAS) y diast&#243;lica (PAD), identificados en la columna de mercurio, sin tener conocimiento de los valores registrados en el aparato autom&#225;tico reportados por otro observador.</p>
      <p>Antes de la evaluaci&#243;n, los individuos permanecieron 5 minutos sentados, en reposo, con el tronco apoyado en una silla y con el brazo relajado. Se calibr&#243; debidamente la columna de mercurio y el aparato autom&#225;tico antes de ejecutarse las afericiones. El tiempo demandado para la evaluaci&#243;n total en cada individuo de la muestra (evaluaci&#243;n antropom&#233;trica y de la presi&#243;n arterial) fue de aproximadamente 10 minutos.</p>
      <p>Al total se realizaron simult&#225;neamente 450 evaluaciones con ambos los instrumentos de medidas de la PA. La clasificaci&#243;n de la variaci&#243;n de los valores de PA, suministrados por el aparato, se dio por las diferencias de los registros de las lecturas entre el man&#243;metro de la columna de mercurio y el aparato autom&#225;tico, en conformidad a los procedimientos descritos por la BHS. El protocolo de la BHS se&#241;ala que los aparatos deben alcanzar al grado B para PAS y PAD, como m&#237;nimo, para que sean instrumentos v&#225;lidos. Los respectivos grados y sus valores est&#225;n descritos en el <xref ref-type="fig" rid="c1-body3">Cuadro 1</xref>.</p>
      <fig id="c1-body3">
        <graphic xlink:href="1678-4170-abc-92-01-0010-00003-gf01-es.jpg"/></fig>
      <p>
        <bold>An&#225;lisis estad&#237;stico</bold>
      </p>
      <p>Inicialmente, se aplic&#243; la prueba de Kolmogorov-Smirnov para verificar la distribuci&#243;n de los datos. Tras ello, se realiz&#243; la estad&#237;stica descriptiva utiliz&#225;ndose valores de promedio y desviaci&#243;n est&#225;ndar (DE), ya que todas las variables analizadas se enmarcaron en el modelo de Gauss de distribuci&#243;n. El an&#225;lisis de variancia para medidas repetidas (con correcci&#243;n de Greenhouse-Geisser cuando violado el presupuesto de la esfericidad) y la prueba <italic>t</italic> de <italic>Student</italic> para muestras pareadas establecieron comparaciones entre los valores promedio de PA encontrados entre los dos aparatos. Para verificar la relaci&#243;n entre los valores de PA suministrados por el aparato autom&#225;tico y la columna de mercurio, se efectu&#243; el coeficiente de correlaci&#243;n intraclase (CCI). Se probaron las medidas de concordancia mediante el m&#233;todo de Bland-Altman (valores num&#233;ricos).Y los par&#225;metros suministrados por la curva ROC (sensibilidad, especificidad y &#225;rea bajo la curva (AUC)) indicaron el poder de diagn&#243;stico de valores elevados de presi&#243;n arterial del aparato oscilom&#233;trico. Se consideraron relevantes, del punto de vista estad&#237;stico, los valores de significancia inferiores al 5%, y los programas utilizados fueron el MedCalc y SPSS versi&#243;n 13.0.</p>
      <p>
        <bold>Resultados</bold>
      </p>
      <p>En la <xref ref-type="table" rid="t1-body3">Tabla 1</xref>, se presentan las caracter&#237;sticas generales de la muestra, agrupadas de acuerdo con el g&#233;nero. No se observaron diferencias estad&#237;sticas entre las variables analizadas.</p>
      <table-wrap id="t1-body3">
        <graphic xlink:href="1678-4170-abc-92-01-0010-00003-gt01-es.jpg"/></table-wrap>
      <p>Los resultados expuestos en la <xref ref-type="table" rid="t2-body3">Tabla 2</xref> demuestran, en valores porcentuales, las diferencias entre las medidas obtenidas por el aparato autom&#225;tico y la columna de mercurio. De acuerdo con los resultados encontrados, el aparato probado obtuvo grado A tanto para la PAS como para la PAD, seg&#250;n los criterios establecidos por la BHS.</p>
      <table-wrap id="t2-body3">
        <graphic xlink:href="1678-4170-abc-92-01-0010-00003-gt02-es.jpg"/></table-wrap>
      <p>En la <xref ref-type="table" rid="es_03t3-body3">Tabla 3</xref> se presentan las comparaciones y los CCI para los valores de PAS y PAD. Entre los aparatos utilizados y en todas las tres afericiones, hubo diferencias significativas para los valores de PAS. Sin embargo, para la PAD hubo diferencia s&#243;lo en la primera aferici&#243;n. En ambos los aparatos, no se observaron diferencias estad&#237;sticas para los valores de PAS y PAD entre las tres afericiones realizadas. Respecto al g&#233;nero, los valores de correlaci&#243;n para PAS, observados entre los aparatos, oscilaron de moderado hacia alto (CCI = 0,74 y CCI = 0,88). Para la PAD, se observaron valores moderados (CCI = 0,50 y CCI = 0,79). En ambas las circunstancias, no se observaron diferencias entre los g&#233;neros que comprometieran la utilizaci&#243;n del equipo.</p>
      <p content-type="asset">
        <table-wrap id="es_03t3-body3">
          <graphic xlink:href="1678-4170-abc-92-01-0010-00003-gt03-es.jpg"/></table-wrap>
      </p>
      <p>Se utiliz&#243; el m&#233;todo de Bland-Altman (<xref ref-type="fig" rid="f1-body3">Fig. 1</xref>) como indicador de concordancia para los valores de PA reportados por ambos los aparatos. Para los valores de PAS, en las 450 afericiones, se verific&#243; que s&#243;lo un 2,2% de las observaciones estaban fuera del intervalo de confianza del 95% estipulado para el promedio de las diferencias observadas. Para la PAD, s&#243;lo el 2,4% de las observaciones estaban fuera del igual intervalo de confianza.</p>
      <fig id="f1-body3">
        <graphic xlink:href="1678-4170-abc-92-01-0010-00003-gf02-es.jpg"/></fig>
      <p>En las 450 afericiones, tras la clasificaci&#243;n de los individuos seg&#250;n su comportamiento hemodin&#225;mico en reposo (normotensi&#243;n/PA elevada), se prob&#243; la eficiencia del aparato autom&#225;tico en diagnosticar la presencia de la PA elevada (<xref ref-type="table" rid="t4-body3">Tab. 4</xref>).</p>
      <table-wrap id="t4-body3">
        <graphic xlink:href="1678-4170-abc-92-01-0010-00003-gt04-es.jpg"/></table-wrap>
      <p>La <xref ref-type="table" rid="t4-body3">Tabla 4</xref> indica que se observaron, en las tres afericiones, elevados valores de sensibilidad (indicaci&#243;n correcta de la presencia de la variable desenlace), de especificidad (indicaci&#243;n correcta de la ausencia de la variable desenlace) y, consecuentemente, de AUC (coeficiente general del diagn&#243;stico, derivado de la interacci&#243;n entre sensibilidad y especificidad).</p>
      <p>
        <bold>Discusi&#243;n</bold>
      </p>
      <p>Este estudio tuvo como prop&#243;sito validar el aparato oscilom&#233;trico autom&#225;tico Omron HEM 742 de medida de PA en ni&#241;os y adolescentes de acuerdo con el protocolo propuesto por la BHS. Tuvo asimismo como objeto evaluar su eficiencia en la clasificaci&#243;n de los individuos de esta franja etaria, en relaci&#243;n a los valores de PA de reposo. Para ello, se tom&#243; en cuenta tanto la validaci&#243;n de este aparato s&#243;lo en individuos adultos<sup>10</sup>, como el hecho de que la mayor&#237;a de los estudios implicando la validaci&#243;n de otros aparatos autom&#225;ticos para la aferici&#243;n de la PA se realiz&#243; en poblaciones adultas<sup>11-14</sup>. Adem&#225;s, con el incremento observado en la ocurrencia de valores elevados de PA de reposo entre ni&#241;os y adolescentes<sup>4,5</sup>, es necesario tambi&#233;n la realizaci&#243;n de estos estudios en poblaciones m&#225;s j&#243;venes, en el sentido de crearse estrategias que faciliten la medida de PA y su interpretaci&#243;n en esta poblaci&#243;n, no s&#243;lo en la pr&#225;ctica cl&#237;nica, sino tambi&#233;n en investigaciones.</p>
      <p>De esa forma, se debe resaltar que aparatos de insuflaci&#243;n y deflaci&#243;n autom&#225;tica son ligeros, port&#225;tiles y de f&#225;cil aplicaci&#243;n e interpretaci&#243;n, reduci&#233;ndose la ocurrencia de errores de lectura por el evaluador, lo que es relativamente observado por el m&#233;todo de auscultaci&#243;n. Por lo tanto, la utilizaci&#243;n de aparatos autom&#225;ticos parece ser una estrategia interesante para la mensuraci&#243;n de la PA, principalmente cuando se trata de un mayor control residencial de los niveles de presi&#243;n por parte de poblaciones hipertensas.</p>
      <p>Para que la validaci&#243;n del uso de un aparato en poblaciones especiales, la BHS<sup>8</sup> sugiere que el aparato sea previamente validado en adultos, como es el caso del aparato Omron HEM 742, validado por Coleman et al<sup>10</sup>; y que se eval&#250;en diferentes valores de PA, en una variaci&#243;n de 100 mmHg hacia 240 mmHg para PAS y de 60 mmHg hacia 120 mmHg para PAD. Ello, sin embargo, fue una limitaci&#243;n del estudio, ya que los valores de PAS oscilaron de 98 mmHg hacia 135 mmHg y de 53 mmHg hacia 86 mmHg para la PAD. Sin embargo, esa variaci&#243;n encontrada en la presente investigaci&#243;n es semejante a la observada en otros estudios de validaci&#243;n de equipos para la mensuraci&#243;n de la PA en poblaciones j&#243;venes<sup>15,16</sup>, en los cuales no se encontraron valores superiores a 140 mmHg para PAS y 90 mmHg para PAD .</p>
      <p>Ello ocurre porque los &#237;ndices de prevalencia de hipertensi&#243;n arterial en esa franja etaria son menores<sup>17,18</sup> cuando comparados a la poblaci&#243;n adulta<sup>18-22</sup>. Adem&#225;s de esto, se necesita resaltar que el diagn&#243;stico de la hipertensi&#243;n arterial en la poblaci&#243;n joven ocurre por medio de porcentiles, considerando el g&#233;nero, la edad y la estatura, diferentemente de los adultos que poseen valores de PAS y PAD preestablecidos.</p>
      <p>En el presente estudio, en todas las tres afericiones realizadas en la muestra, los valores de PAS suministrados por el equipo probado sobreestimaron significativamente los valores obtenidos por el instrumento de columna de mercurio. Esos mayores valores suministrados por el aparato oscilom&#233;trico pueden provenir de medidas m&#225;s bajas producidas por el evaluador del m&#233;todo de auscultaci&#243;n y/o ser realmente valores m&#225;s altos producidos por el aparato prueba<sup>23</sup>. Pero, una explicaci&#243;n definitiva para este comportamiento no est&#225; totalmente definida en la literatura. En relaci&#243;n a la PAD, s&#243;lo se encontr&#243; diferencia significativa en la primera aferici&#243;n. A pesar de dichas diferencias, los valores de PAS y PAD presentados por el aparato Omron HEM 742 y por equipo de columna de mercurio no oscilaron de manera significativa entre las tres afericiones, cuando comparados consigo mismos, lo que proporciona una buena reproductibilidad de los datos en ambos los instrumentos.</p>
      <p>El CCI se present&#243; de manera significativa con valores moderados/elevados en todas las afericiones, tanto para PAS como para PAD, resultando en una buena relaci&#243;n entre los valores observados para ambos los instrumentos. De este modo, cuanto mayor sea el valor reportado por un equipo, mayor ser&#225; el valor encontrado por otro aparato para el mismo individuo. Pero, tal procedimiento estad&#237;stico es s&#243;lo una medida de relaci&#243;n: no indica la concordancia entre esos equipos y, consecuentemente, no detecta sus posibles variaciones extremas. Dicha limitaci&#243;n es suplantada por la utilizaci&#243;n del m&#233;todo de Bland-Altman que, a su vez, indica la concordancia entre datos num&#233;ricos. Se observ&#243;, en el presente estudio, una baja discordancia entre los valores de PA encontrados por ambos los aparatos. Este hecho se confirm&#243; cuando se alcanzaron los criterios propuestos por la BSH<sup>8</sup>, atingi&#233;ndose, incluso, la gradaci&#243;n m&#225;s elevada.</p>
      <p>Dichos indicativos positivos presentados por el CCI y por el m&#233;todo de Bland-Altman son informaciones importantes sobre la confiabilidad de la utilizaci&#243;n del instrumento. Sin embargo, se tratan de valores num&#233;ricos y, por ello, no son capaces de indicar la eficiencia del instrumento en la pr&#225;ctica cl&#237;nica, de manera m&#225;s espec&#237;fica en la detecci&#243;n de valores elevados de PA, siendo necesario emplear la curva ROC para este fin. As&#237;, los par&#225;metros suministrados por la curva ROC indicaron que el aparato Omron HEM 742 present&#243; elevada capacidad de detectar tanto la presencia (sensibilidad) como la ausencia (especificidad) de valores elevados de PA en la muestra investigada.</p>
      <p>
        <bold>Conclusi&#243;n</bold>
      </p>
      <p>El aparato Omron HEM 742 presenta buena confiabilidad para uso en la evaluaci&#243;n de adolescentes, una vez que se lo valid&#243; de acuerdo con las recomendaciones internacionales de la BHS<sup>8</sup>, alcanzado as&#237; el grado A para los an&#225;lisis de la PAS y PAD, adem&#225;s de haber presentado buena sensibilidad y especificidad en el diagn&#243;stico en relaci&#243;n a los niveles de presi&#243;n de individuos de esa franja etaria.</p>
      <p>
        <bold>Potencial Conflicto de Intereses</bold>
      </p>
      <p>Declaro no haber conflicto de intereses pertinentes.</p>
      <p>
        <bold>Fuentes de Financiaci&#243;n</bold>
      </p>
      <p>El presente estudio no tuvo fuentes de financiaci&#243;n externas.</p>
      <p>
        <bold>Vinculaci&#243;n Acad&#233;mica</bold>
      </p>
      <p>Este art&#237;culo es parte de disertaci&#243;n de Maestr&#237;a de Diego Giulliano Destro Christofaro, por la Universidad Estatal de Londrina.</p>
      <p>
        <bold>Referencias</bold>
      </p>
    </body>
  </sub-article>
</article>